background image

8

the tip at about one inch per second 
along seams or joints suspected of 
leaking. When the probe encoun-
ters a leak, both the audio pitch and 
the flashing rate of the neon lamp 
will increase. Both will continue to 
respond at the faster rate as long as 
the probe is held at the leak.

9. For detecting much larger or smaller 

leaks, set the leak size switch to the 
appropriate range LG or MED,  readjust 
the balance control for slow ticking 
(1–3 ticks per second), and probe 
for leaks as in the previous step. The 
calibrate reference cannot be used to 
calibrate these two ranges. The factory, 
however, can supply leak standards 
of any value required for verification of 
capability to specific leak rates.

Sensor Life Extension

  The H-10G provides a sensor adjust-
ment that substantially extends sensor 
life. This procedure also provides for a 
more accurately calibrated sensitivity 
throughout the sensor life, and allows 
compensation for abnormally high or low 
line voltages (see 

Maintenance 

section).

  The sensor heat control is a screw-
driver adjustment accessible through a 
hole on the bottom of the case. As the 
H-10G is used and the sensor ages, its 
sensitivity will gradually decrease. 
  The overall sensitivity and calibra-
tion of the H-10G may be restored by 
gradually adjusting the sensor heat 
control clockwise until finally, with a fully 
aged sensor, the control knob will be 
turned clockwise to indicate 2 o’clock 
(fig. 4). After that, when the sensor 
fails to respond properly, replace the 
sensor following the procedure in the 

Maintenance 

section and readjust the 

sensor heat control back to 7 o’clock 

Summary of Contents for H-10G

Page 1: ...performance service leak detector available in the indus try today It will detect refriger ants with a sensitivity that will effectively pinpoint both small and large leaks quickly and efficiently We...

Page 2: ...scope of its specifications For details on appropriate use refer to the gen eral model description application and operation discussions in this manual Before risking equipment damage or personal inju...

Page 3: ...nse to an industry that is in the process of converting to more environmentally safe refrigerants worldwide Ozone depleting chlorine is being reduced or eliminated in most refrigerants To adequately l...

Page 4: ...he H 10G for the first time remove the paper disc from the cap on the calibrate reference bottle fig 1 and allow a few minutes for the leak rate to stabilize This disc is used to avoid evaporation and...

Page 5: ...e that the actual height and final resting position of the red ball are not important See Figure 2 If the airflow ball does not rise Tap the probe lightly to ensure the ball is not sticking Check the...

Page 6: ...calibrate reference bottle for 2 3 seconds This allows enough time for the H 10G to pump down the concentration of gas which normally collects around the bottle s outlet and thus arrive at a steady st...

Page 7: ...rine based refrigerants Please refer to the Product Sensitivity Levels chart on the pocket of the inside lid or see this manual s Specifications section The term R12 sensitivity is used to represent c...

Page 8: ...nsor Life Extension The H 10G provides a sensor adjust ment that substantially extends sensor life This procedure also provides for a more accurately calibrated sensitivity throughout the sensor life...

Page 9: ...ery little maintenance A kit that contains the necessary expendable replacement parts is supplied with each unit Instructions for replacing these items are described in the following sec tions Replaci...

Page 10: ...the cover 5 Set the sensor heat control fig 4 to its 7 o clock position 6 Place the unit in operation and check its sensitivity by following the procedure in steps 5 6 and 7 in the Operating Instruct...

Page 11: ...arger leak size Problem Excessive response to cali brate reference Cause High line voltage Solution Reset sensor heat control more counterclockwise Problem Audio normal but probe lamp does not flash C...

Page 12: ...sor exposed to excessive amounts of halogen 4Water in probe 5Low line voltage Solution 1Replace with part 3015 0864 2Replace filter in probe tip Check operation of pump 3Place probe to clean air and a...

Page 13: ...e no movement Switch Position LG MD SM R134a 5 0 0 5 0 1 R12 3 0 0 1 05 Power Supply 10 100 120 220 240 volts 50 60 Hz 20 watts in accor dance with rating plate Response Time Approx 1 second Reference...

Page 14: ...3015 0641 Clear Probe Tip 3015 0680 Rubber Probe Tip 3015 0486 Replacement Sensor 3015 0737 Maintenance Kit 12 Filters 4 Airflow Balls 3015 0781 Tune Up Kit Sensor 100 Filters 4 Airflow Balls 3 Rubber...

Page 15: ...al fee Please include your company name telephone number includ ing area code and a contact name Limited Warranty The purchaser is warranted that this leak detector will be free of defects in material...

Page 16: ...ce Center 621 Hunt Valley Circle New Kensington PA 15068 Phone 724 334 5051 Fax 724 334 5723 E mail help MyBacharach com Canada Bacharach of Canada Inc 20 Amber St Unit 7 Markham Ontario L3R 5P4 Canad...

Page 17: ...17 Notes...

Page 18: ...18 Notes...

Page 19: ...19 Notes...

Page 20: ...3015 0846 REV 8 9 2012 621 Hunt Valley Circle New Kensington PA 15068 Ph 724 334 5000 Fax 724 334 5001 Toll Free 1 800 736 4666 Website http www MyBacharach com E mail help MyBacharach com...

Reviews: