Maintenance

Probe

The probe’s sponge-type 

fi

 lter (Figure 3, Item 9) needs to be in spect ed at 

reg u lar intervals for dirt contamination. Replace 

fi

 lter as nec es sary by 

fi

 rst pulling out the Peltier cooler from the probe and then ex chang ing 

the 

fi

 lter with a new one.

Peltier Cooler

The cooler’s wadding 

fi

 lter (Figure 3, Item 12) needs to be in spect ed at 

regular in ter vals for dirt contamination. Re place 

fi

 lter as nec es sary by 

fi

 rst pulling off the 

fi

 lter chamber’s top cap and then ex chang ing the 

fi

 lter with a new one.

Condensate Pump

WARNING!

 

 The condensed water will be warm and slightly 

acid ic. Take all necessary precautions to pre vent the water 

from splashing into your eyes or con tact ing your skin.

Empty the pump’s reservoir (Figure 3, Item 15) by 

fi

 rst holding the 

pump with its reservoir facing downwards. Then twist the reservoir 90 

de grees (either clock wise or counterclockwise) to free it from the pump 

base. Empty the res er voir; then reattach it to the pump. Before re at -

tach ing the reservoir, however, make sure that the res er voir’s o-ring 

(Figure 3, Item 20) is prop er ly po si tioned inside its grove in the pump 

base.

The pump’s wadding 

fi

 lter (Figure 3, Item 17) needs to be in spect ed at 

regular in ter vals for dirt contamination. Replace 

fi

 lter as nec es sary by 

fi

 rst unscrewing the 

fi

 lter chamber cap and then ex chang ing the 

fi

 lter 

with a new one.

Summary of Contents for Flue-Gas Sample Conditioning System

Page 1: ...nd SO2 Water vapor is removed from the flue gas sample by passing the sample through a Peltier Cooler where the sample is chilled which in turn causes the water vapor contained in the gas sample to be removed from the sample The water that was extracted from the gas sample is then pumped out of the Peltier Cooler and into a water trap The now dry conditioned flue gas sample is then sent to its ass...

Page 2: ...and then inserting the cooler s inlet connector ensuring first that a filter is present between the cooler and probe Attach the pump s inlet hose to the bottom of the Peltier cooler Water that is condensed from the flue gas sample will be pumped into the con densate pump s reservoir through this hose Note that the condensate pump can be placed in an upright horizontal or hanging position A magnet ...

Page 3: ... System 4 Bacharach Inc Instruction 0024 9415 Figure 2 Equipment Setup PELTIER COOLER PROBE ASSEMBLY PUMP POWER SUPPLY 100 240 VAC 50 60 HZ Connect to Combustion Test Equipment STACK TEMP THERMOCOUPLE PRESSURE Blue FLUE GAS Yellow ...

Page 4: ...alyzer as described in its in struction manual Then insert the probe into the flue stack and begin testing During the combustion test the condensed water vapor from the flue gas will accumulate inside the pump s reservoir If the water level be gins to approach the ends of the tubes inside the reservoir then stop the test and empty the reservoir as described in the Maintenance Section of this manua...

Page 5: ...ARNING The condensed water will be warm and slightly acidic Take all necessary precautions to prevent the water from splashing into your eyes or contacting your skin Empty the pump s reservoir Figure 3 Item 15 by first holding the pump with its reservoir facing downwards Then twist the reservoir 90 degrees either clockwise or counterclockwise to free it from the pump base Empty the reservoir then ...

Page 6: ... 0875 7 Screw M4 x 6mm 0501 3829 8 Retaining Nut 0102 3740 9 Filter Qty 5 0007 1646 Peltier Cooler Power Supply 10 Complete Assembly Peltier Cooler Power Supply AC Power Cord not included 0024 1179 11 Plastic End Cap 0106 5454 12 Filter Wadding 0007 1647 13 Power Supply 0304 1713 14 AC Power Supply Cord 3 Wire 4998 8986 Condensate Pump 15 Complete Assembly 0024 1180 16 Reservoir 0106 5453 17 Plast...

Page 7: ...Flue Gas Sample Conditioning System 8 Bacharach Inc Instruction 0024 9415 Figure 3 Replacement Parts 11 20 20 12 Part of Item 10 1 21 9 6 7 8 5 2 Part of Item 10 13 14 18 23 17 22 16 20 19 15 ...

Page 8: ...cting a Bacharach Service Center at the following locations United States Bacharach Inc 621 Hunt Valley Circle New Kensington PA 15068 Phone 724 334 5051 Fax 724 334 5723 Email help mybacharach com Canada Bacharach of Canada Inc 20 Amber St Unit 7 Markham Ontario L3R SP4 Canada Phone 905 470 8985 Fax 905 470 8963 Email bachcan idirect com ...

Page 9: ...Flue Gas Sample Conditioning System 10 Bacharach Inc Instruction 0024 9415 ...

Page 10: ...Bacharach Inc 11 Flue Gas Sample Conditioning System Headquarters 621 Hunt Valley Circle New Kensington PA 15068 Ph 724 334 5000 Fax 724 334 5001 Toll Free 800 736 4666 Website www mybacharach com E mail help mybacharach com ...

Reviews: