![Bacharach Flue-Gas Sample Conditioning System Instruction Download Page 2](http://html1.mh-extra.com/html/bacharach/flue-gas-sample-conditioning-system/flue-gas-sample-conditioning-system_instruction_3366362002.webp)
Instruction 0024-9415
Bacharach, Inc.
3
Flue Gas Sample Conditioning System
Technical Data
Probe:
Standard Length ......................................................... 11.5 in (295 mm)
Optional Lengths ............................................................. 7 in (180 mm)
19.5 in (500 mm)
39 in (1000 mm)
Peltier Cooler:
Inlet Sample Dew Point.......................................... 149 °F (65 °C) max.
Outlet Sample Dew Point .................18 °F (10 °C) min. below ambient
Cooling Power ................................................................................. 10 W
Environment Temperature .............................32 to 104 °F (0 to 40 °C)
Weight ................................................................................8.5 oz (240 g)
Condensate Pump:
Sample Flow Rate ..........................................................1.5 LPM @ STP
Length of Hose ...................................................................... 6.6 ft (2 m)
Reservoir
Storage
........................................................................... 50 ml
Environment Temperature .............................32 to 104 °F (0 to 40 °C)
Weight ..............................................................................12.7 oz (360 g)
Power Supply:
Input ..............................................................100 to 240 VAC, 50/60 Hz
Output ............................................................ 9 VDC, max. @ 4 A, max.
Equipment Setup
Refer to Figure 2. If not already done, attach the Peltier cooler to the
probe as sem bly by
fi
rst removing the plastic end cap from the rear of
the probe, and then inserting the cooler’s inlet connector, en sur ing
fi
rst
that a
fi
lter is present between the cooler and probe.
Attach the pump’s inlet hose to the bottom of the Peltier cooler. Water
that is condensed from the
fl
ue-gas sample will be pumped into the con-
densate pump’s reservoir through this hose. Note that the con den sate
pump can be placed in an upright, horizontal, or hanging po si tion. A
mag net mounted in the base of the pump allows the pump to be at-
tached to overhead ductwork.
Connect the power supply’s DC output connector to the con den sate
pump; then connect the condensate pump’s DC output power con nec tor
to the Peltier cooler. The Peltier cooler and its associated con den sate
pump are started by plugging the power supply into a source of AC
power that is be tween 100 and 240 VAC, 50/60 Hz.
Connect the probe assembly’s stack-temperature thermocouple con-
nec tor, pres sure
fi
tting, and
fl
ue-gas
fi
tting to the combustion an a lyz er
per the analyzer’s instruction manual.