background image

Single Gas Instrument Service Manual

Instruction 0019-9166 

Bacharach, Inc.

 Page 

9

 2.4 

Sensor 

Replacement

When the instrument can no longer be spanned, as described in Section “3.0 Sen-

sor Calibration,” the Sensor has reached the end of its useful life, and should be 

replaced. The life of a Sensor is typically two years. The Sensor may also need 

to be replaced if the instrument will not zero properly, or if erratic gas readings 

occur. Refer to Section “4.0 Troubleshooting Chart.”

Equipment Needed (Refer to Section “5.0 Replacement Parts” for Part 

Nos.):  

  Phillips-head screwdriver, No. 1 

 Replacement Sensor:

  Nutdriver, 4mm (5/32") 

  Dioxor II - SO

2

 

 Screwdriver, 1/8" 

fl

 at blade 

  Monoxor II - CO

  O-Ring and Membrane, if required 

  Nonoxor II - NO

X

 

 

 

  Monoxor II H - CO

Procedure:  

WARNING!

 

Burn hazard! The Sensors contain corrosive chemicals, 

 

DO NOT puncture or take the sensors apart.

1.  Open the case as described in Section “2.1 Opening the Case.”

2.  Using a nutdriver, loosen the nuts that secure the sensor wires to the Sen-

sor terminals R, S and C; then remove the wires. If the Sensor has pin and 

socket connections, simply pull the sockets from the sensor pins.

3. Using 

fl

 at-blade screwdriver, remove the three screws and nuts that secure 

the Sensor to the chamber assembly; then remove the Sensor along with its 

O-ring and Membrane from the chamber assembly.

4.  Remove and discard of the old Sensor in accordance with local and federal 

regulations. DO NOT discard the sensor’s O-ring and Membrane, unless you 

intend to replace these items.

NOTE:

 Replace the O-ring if it looks deformed, and replace the 

Membrane if it is contaminated with oil or dirt. We recommend 

that both the O-ring and Membrane be replaced at the same time.

5.  Remove the new Sensor from its shipping canister. Do not remove the short-

ing wire (across the R & S terminals) from the Sensor (SO

2

 & CO only, the 

NO

X

 does not have a shorting wire) until you are ready to make the electrical 

connection.

6.  Position the Membrane, O-ring, and new Sensor on the chamber assembly 

as per Fig 2-2. Ensure that the sensor terminals are positioned as shown. 

Secure these items to the chamber assembly using the screws and nuts re-

moved in Step 3 (Torque to 4–8 oz. in.). 

DO NOT overtighten the screws!

7.  Remove the shorting wire from the R & S terminals (SO

2

 & CO only). Con-

nect the 

blue

 (S), 

red

 (R), and 

black

 (C) wires from the P.C.B. to their as-

sociated sensor terminals (See Fig 2-4). 

If the Sensor uses nuts to secure the 

Sensor wires, DO NOT overtighten these connections!

8.  Close the case as per Section “2.2 Closing the Case.”
9.  Calibrate the new Sensor as per Section “3.0 Sensor Calibration.”

Summary of Contents for DIOXOR II

Page 1: ...ERVICE MANUAL for DIOXOR II MONOXOR II MONOXOR II H High Range NONOXOR II Rev 7 May 2010 Bacharach Inc 621 Hunt Valley Circle New Kensington PA 15068 Phone 1 800 736 4666 Fax 724 334 5001 Web www myba...

Page 2: ...Bacharach Dioxor Monoxor and Nonoxor are registered trademarks of Bacharach Inc Copyright Bacharach 1997 2010 all rights reserved A...

Page 3: ...it Board Replacement 10 2 6 ChamberAssembly Servicing 11 3 0 SENSOR CALIBRATION 12 3 1 Monoxer II CO Calibration 12 3 2 Dioxor II SO2 Monoxor II H CO and Nonoxor II NOx Calibration 13 4 0 TROUBLESHOOT...

Page 4: ...he same in this instruction Instrument maintenance which includes information on how to Open and close the case Clean the Fan Assembly and chamber assembly area Replace the sensor Replace the Printed...

Page 5: ...c strap over top of the instrument Fig 2 1 2 Slide off battery cover 3 Remove batteries 4 Using a No 1 Phillips head screwdriver remove the two screws from the rear of the case 5 Lift up the bottom po...

Page 6: ...r the chamber assembly is in place 1 Ensure that the P C B is properly positioned face down over its locating tab on the front case Fig 2 2 2 Properly position the chamber assembly between the ribs of...

Page 7: ...ing the two screws and nuts that secure the Chamber to the Motor Assembly Fig 2 3 NOTE Do not remove the Fan Assembly unless it s absolutely neces sary for cleaning or repair purposes 4 If it is neces...

Page 8: ...ng fully Tighten the set screw to approxi mately 32 40 ounce inches using the 05 in Allen wrench Spin the Fan Assembly to be sure it will run in the chamber b Mate and secure the Chamber and Motor Ass...

Page 9: ...hat secure the Sensor to the chamber assembly then remove the Sensor along with its O ring and Membrane from the chamber assembly 4 Remove and discard of the old Sensor in accordance with local and fe...

Page 10: ...ensor connector from the P C B NOTE For SO2 CO Sensors only If the Sensor is not going to be immediately plugged into the new P C B place a shorting wire between sensor terminals R and S 4 Remove the...

Page 11: ...he Sensor along with its O ring and Membrane from the chamber assembly 3 Using a No 1 Phillips head screwdriver disassemble the Chamber from the Motor Assembly Fig 2 3 4 To remove the Fan Assembly ref...

Page 12: ...edure WARNING Perform this procedure in a well ventilated area CO is toxic 1 Assemble the calibration equipment by connecting the Carbon Monoxide Gas Cylinder and the components of the Calibration Kit...

Page 13: ...nel POWER switch to OFF 8 Disassemble the calibration equipment setup and store the components in the calibration kit s carrying case The presence of SO2 and NOX gas in small amounts and CO gas in lar...

Page 14: ...te be fore switching on unit to allow its PTC resistor to cool Replace P C B Refer to Section 2 5 Printed Circuit Board Replacement Replace P C B Refer to Section 2 5 Printed Circuit Board Replacement...

Page 15: ...defective P C B defective Replace batteries with four fresh C Alkaline cells Break in a new Motor for at 30 minutes Replace Motor Assembly Refer to Section 2 6 Chamber Assembly Servicing Check Sensor...

Page 16: ...ns See Fig 2 2 Run Motor for at least 3 hours to wear away arc spots If display still bounces replace Motor Assembly Refer to Sec tion 2 6 Chamber Assembly Servicing Tighten screws Refer to Sec tion 2...

Page 17: ...not shown 0019 3104 2 Probe Tube 0019 3073 3 Probe Handle 0019 3017 4 Screw Phillips head M3 x 5 x 8 mm 0501 3806 5 Tubing 5 16 OD X 3 16 ID 0003 6109 6 Filter Saturator Bushing 0011 0074 7 Filter Pac...

Page 18: ...1 O Ring and Membrane 0024 0520 12 Printed Circuit Board Assy Dioxor II 0019 3022 Printed Circuit Board Assy Monoxor II 0019 3012 Printed Circuit Board Assy Nonoxor II 0019 3020 Printed Circuit Board...

Page 19: ...Single Gas Instrument Service Manual Instruction 0019 9166 Bacharach Inc Page 19 Figure 5 3 Replacement Parts Back and Front Exploded Views...

Page 20: ...Monoxor II H 0019 9180 Calibration Card 0019 9150 Service Manual This Manual 0019 9166 Label Carrying Case 0019 3077 Probe Stop for Securing Probe in a Flue or Stack 0024 0434 Warranty Card 0006 8825...

Reviews: