If batteries are accidently installed in the wrong polarity, a positive
temperature coefficient (PTC) thermistor will protect the instrument’s
electronic circuity. The instrument will operate once batteries are
properly installed and the PTC thermistor is allowed to cool.

3.2  Probe Installation

The probe comes disassembled. To add the probe tip:

a. Loosen but DO NOT REMOVE the Knurled Nut (03-2023) on

the main part of the Probe Assembly (19-3076).

b. Insert probe tube into Extended Adapter Block (19-3072).
c. Hand tighten the Knurled Nut (03-2023) then add 1/4 turn to

snug probe to the rest of the probe assembly.

NOTE: In order to prevent the possibility of misplacing the
Front (03-2024) and Back (03-2025) Ferrules, loosen but
DO NOT REMOVE the Knurled Nut (03-2023) before
 inserting or removing the Probe Tube (19-3073).

Install the probe by sliding the end of its tubing over the  gas inlet port
on the top right side of the instrument as shown in Figure 3. The tubing
may be difficult to slide over the gas inlet port of the unit for the first
time. This was done intentionally to allow for a snug fit. Use a little dish
washing liquid diluted in water or heat the end of the tube in hot tap
water to help it slide onto the port.

Figure 3. Probe Installation

3

Summary of Contents for DIOXOR 19-7038

Page 1: ...9 9121 Portable SO2 Analyzer Part No 19 7038 19 7043 Rev 5 September 1997 Bacharach Inc 625 Alpha Drive Pittsburgh PA 15238 412 963 2157 2164 Fax 412 963 2640 Web www bacharach inc com Printed in U S...

Page 2: ...such as electrochemical sensors are warranted for a period of six months THE FOREGOING WARRANTIES ARE EXCLUSIVE AND ARE GIVEN AND ACCEPTED IN LIEU OF I ANY AND ALL OTHER WARRANTIES EXPRESS ORIMPLIED...

Page 3: ...be obtained before use Assistance is available by calling your nearest Bacharach Service Center 1 INTRODUCTION The Dioxor II is a commercial grade portable instrument designed to display concentratio...

Page 4: ...escribed in the following paragraphs For your convenience and to ensure that the instrument will provide reliable SO2 indica tions the SO2 sensor is installed and the instru ment calibrated on a known...

Page 5: ...n the Knurled Nut 03 2023 then add 1 4 turn to snug probe to the rest of the probe assembly NOTE In order to prevent the possibility of misplacing the Front 03 2024 and Back 03 2025 Ferrules loosen bu...

Page 6: ...ove the batteries and ensure that the POWER switch is OFF The OFF position places a short across the SO2 sensor thus keeping the sensor from being destabilized while not in use 4 2 Zeroing the Instrum...

Page 7: ...of 025 ppm 4 3 Backlight ON OFF The LCD can be read in low light areas by setting the front panel LIGHT switch to ON The backlight stays on until turned off or until the POWER switch is set to OFF 4 4...

Page 8: ...conditions 4 7 Long Term Storage When storing the Dioxor II for extended periods of time set its POWER switch to OFF and remove the batteries The POWER OFF positionplacesashortacrosstheSO2 sensor thu...

Page 9: ...Bacharach S S Center 7300 Industrial Park Rt 130 Bldg 22 Pennsauken NJ 08110 609 665 6176 609 665 6661 FAX Bacharach S S Center 5151 Mitchelldale B 4 Houston TX 77092 713 683 8141 713 683 9437 FAX 7...

Page 10: ...ion of the nose and throat choking sneezing coughing and broncho spasm Severe overexposure may cause death through pulmonary edema abnormalfluidbuildupinthelungs orfromrespiratoryarrest Prolonged or r...

Page 11: ...cult Seek medical attention immediately References 1 Code of Federal Regulations Title 29 CFR Parts 1900 1910 Labor Superintendent of Documents U S Government Printing Office Washington DC 20402 2 Thr...

Reviews: