Instruction 3015-4286
Safety Precautions
AC Power Supply
The AGM-SZ uses a universal power supply that is capable of accepting inputs of 100 to 240 VAC, 50/60 Hz.
The monitor’s power consumption is 15 Watts. It is highly suggested that the monitor be connected directly to
the AC power source, preferably on its own circuit (with UPS or surge protection).
A switch or circuit breaker rated 1.0 A, 250 VAC, with a minimum terminal spacing of 3.0 mm must be attached
to the monitor’s AC power leads. This switch must also be located in close proximity to the monitor, and be in
easy reach of the operator. This switch should also be clearly marked as the monitor’s main AC disconnect
device.
Protective Grounding
Under no circumstances should the AGM-SZ be operated without connection to a protective ground. Doing so
poses a potential shock hazard and is also a violation of electrical safety standards applicable to this type of
equipment.
Explosive Atmosphere
Do not operate this equipment in the presence of flammable liquids, vapors or aerosols. Operation of any
electrical equipment in such an environment constitutes a safety hazard.
Proper Exhaust Venting
It is imperative that the monitor’s exhaust port be vented as described in this manual. Failure to do so may
constitute a safety hazard.
Accessing the Interior of the Monitor
Extreme care should be exercised when accessing the interior of the monitor. Only qualified electrical
maintenance personnel should perform connections and adjustments. Always remove AC power before
performing work inside the monitor.
Misuse & Modifications to the Monitor
The protection provided by the monitor may be impaired if the monitor is used in a manner not specified by
Bacharach, Inc. Changes or modifications to this monitor, not expressly approved, will void the warranty.
In Case of Malfunction
Do not continue to use this equipment if there are any symptoms of malfunction or failure. In the case of such
occurrence, de-energize the power supply and contact a qualified repair technician or the nearest Bacharach
Service Center. Use ONLY the provided knockouts for electrical and communication wiring. Drilling into the box
will void the warranty.
Fusing
F1, F2:
1.0 A, 250 V, Type “F”
Installation Category
Installation Category II, Pollution Degree II, as defined by UL.
Altitude Limit
6,562 ft (2,000 m)
Cleaning
To clean the outside of the case,
DO NOT
use soap and water.
USE
a dry cloth.