The following screens show the “Full Level Trim” or two-point calibration.
Full Level Trim
Two-point calibration requires actual level simulation, meaning the transmitter needs to detect level at
the 0 and 100% points.
There is almost never a need to do a Full Level Trim, the LTM-350 (and 250) has 1-point calibration
capabilities which allow calibration with existing level conditions. Also there are factory default
configurations can be utilized to revert the transmitter back to its original settings (at time of shipment).
39