The following screens show the D/A Trim method and Loop Test Method.
D/A Trim
31
The “Loop Current” menu has the parameters that affect
the analog output of the transmitter. The Alarm Type can
be change from high to low or holdout. D/A trim is used
to scale the accuracy of 4.00 mA and 20.00 mA. Loop Test
is very useful in startup situation where the transmitters
operation can be verified and the 4/20 loop can be
checked with control system. The D/A trim and Loop Test
are methods that go through a step by step procedure.
Summary of Contents for LTM Series
Page 35: ...32 ...
Page 36: ...33 ...
Page 37: ...34 ...
Page 38: ...Loop Test 35 ...
Page 39: ...36 ...
Page 40: ...37 ...
Page 41: ...38 ...
Page 43: ...40 ...
Page 44: ...41 ...
Page 45: ...42 ...
Page 50: ...Operation status lets the user know how the transmitter is operating internally 47 ...
Page 51: ...Calibration Status lets the user know if the instrument has been incorrectly calibrated 48 ...