5. MAINTENANCE AND TROUBLESHOOTING 

 
No routine maintenance is required other than keeping the interior of the unit clean and free of dirt, dust and 

 

other contaminants. The LS7000 consists of two main sub-assemblies. These are the enclosure with the 

 

probe assembly and the power supply board. The following troubleshooting guide will assist in determining 

 

how to correct most of the problems which may occur in the field. Review the Installation and Calibration 

 

procedures in sections 3 & 4. 
 

PROBLEM 

 

 

POSSIBLE CAUSE 

 

 

SOLUTION 

RED LIGHT cannot be   

Antenna lead not connected. 

 

Plug antenna lead into probe. 

adjusted to turn on. 

 

No power to unit. 

 

 

Check for correct power to unit. 

 

 

 

 

Blown on-board fuse.   

 

Replace fuse. 

 

 

 

 

Bad sensing card. 

 

 

Repair or replace card. 

 
RED LIGHT remains on at 

Antenna probe is shorted to case or  

Unplug banana plug and position 

all times 

 

 

ground.  

 

 

 

it so that the bare end is not 

 

 

 

 

 

 

 

 

 

touching anything. Turn calibration 

 

 

 

 

 

 

 

 

 

pot 20 turns CCW. LED should go 

 

 

 

 

 

 

 

 

 

out. If it does, repair, replace or clean 

 

 

 

 

 

 

 

 

 

antenna probe. If LED does not go  

 

 

 

 

 

 

 

 

 

out, replace electronics. 

 

 

 

 

Bad sensing card. 

 

             Repair or replace card. 

  
Unit triggers when material 

Improper mount of probe.            

Contact factory or distributor. 

touches probe, but will not  

Improper calibration procedure.   

See Section 4.4 

reset when material recedes        Excessive material build-up on  probe.  Perform “dirty probe” calibration; 
from probe or unit gives  

 

                                                    recalibrate with buildup material on 

false alarm. 

 

 

 

 

 

                           probe – see section 4.4 

                                                    Probe is mounted in flow material.          Revise mounting or use time delay  
                                                                                                                       
Unit will not detect material. 

Improper calibration. 

 

 

See calibration instructions section 4.

 

 

 

 

 

Antenna lead not plugged into   

Plug antenna lead into probe. 

 

 

 

 

probe.   

 

 

 

 

 

Unit was calibrated with material 

Be sure material is not touching 

 

 

 

 

touching probe.  

  

 

probe and recalibrate. 

 
Unit will not stay in  

 

Poor grounding of unit to vessel. 

Provide secure grounding  

calibration. 

 

 

 

 

 

 

 

connection. 

 

 
Relay operates properly, but 

Burned or broken lands on the    

Turn off power. Remove 

no signal at terminals.   

printed circuit board. 

 

 

module from housing and inspect 

 

 

 

 

 

 

 

 

 

lands  on  printed  circuit  for  damage.

 

 

 

 

 

 

 

 

 

 

Replace if necessary. 

 

 

 

 

Bad relay contacts. 

 

 

Replace  relay  or  return  for  repair.

 

 

Summary of Contents for LS7000

Page 1: ...LS7000 LEVEL SWITCH OWNERS MANUAL INSTALLATION CALIBRATION TROUBLESHOOTING WARRANTY P O Box 70094 Houston TX 77270 713 467 4438...

Page 2: ...r problems arise during installation please contact Customer Service 1 800 835 8012 or 713 467 4438 8 00 a m to 5 00 p m CST US Failure to follow safe installation and servicing guidelines could resul...

Page 3: ...tion 5 1 4 GA Drawing 5 3 INSTALLATION 6 3 1 Inspection 6 3 2 Bench Test 6 3 3 Physical Installation 6 3 4 Removing the Electronics 7 3 5 Typical Applications and Wiring Diagrams 8 4 CALIBRATION 14 4...

Page 4: ...rical Power 115 VAC 15 50 60 Hz 2 watts standard 12 VDC 24 VDC or 230 VAC optional Output 2 Form C contacts DPDT relay 5 amp resistive At 125 250 VAC 30 VDC Fuse On board 250 mA 125 volts Fail Safe Se...

Page 5: ...www babbittinternational com Reference Manual 2021 5 1 3 Ordering Information 1 4 GA Drawing...

Page 6: ...2 On the bench hook up the appropriate power to L1 and L2 The operating voltage of the LS7000 is marked on the power supply board Hold the unit so that nothing is touching any surface and press the t...

Page 7: ...dry materials mount the LS7000 in the top of the vessel whenever possible This allows you to lengthen the probe if necessary b Avoid mounting the LS7000 near the product inlet vent return lines dust c...

Page 8: ...ONTACTS SUPPLY VOLTAGE NOTE EACH LS7000 POWER SUPPLY IS INTENDED FOR ONLY ONE SUPPLY VOLTAGE THIS VOLTAGE IS PRINTED ON THE POWER SUPPLY BOARD FOR 115 VAC OR 230 VAC UNITS HOOK UP THE HOT LEAD TO L1 A...

Page 9: ...www babbittinternational com Reference Manual 2021 9...

Page 10: ...www babbittinternational com Reference Manual 2021 10...

Page 11: ...www babbittinternational com Reference Manual 2021 11...

Page 12: ...www babbittinternational com Reference Manual 2021 12...

Page 13: ...www babbittinternational com Reference Manual 2021 13...

Page 14: ...E S3 and S4 also select fail safe high level FSH or fail safe low level FSL 2 Select the timing range using S1 and S2 as shown on the chart below next page 3 Adjust the time adjustment potentiometer c...

Page 15: ...djustment potentiometer clockwise CW until it comes on 3 Turn the sensitivity adjustment pot counter clockwise CCW until the Red LED just goes out This sets the unit to its maximum sensitivity If in y...

Page 16: ...go out If it does repair replace or clean antenna probe If LED does not go out replace electronics Bad sensing card Repair or replace card Unit triggers when material Improper mount of probe Contact...

Page 17: ...applied to any components which have been damaged by improper installation use exposure to unusual atmospheric conditions or components which have been misused abused damaged by neglect or accident Th...

Reviews: