background image

C.  Calibrate the High Setpoint 

As a rule of thumb, clockwise (CW) rotation of the pot lowers the setpoint; counter-clockwise (CCW) rotation 
of the potentiometer raises the setpoint. 
1)  Raise the fluid level to where you want the high setpoint to be on the probe. 
2)  Observe the high red LED.  If it is off, go to step 4. 
3)  If the high red LED is on, turn the high adjustment pot CCW until the high red LED goes out. 

4) 

Turn the high adjustment pot CW until the high red LED just comes on.  This adjustment must be made 
very carefully.

 

NOTE:

 

Always fill and empty the vessel to verify the accuracy of the calibration.  Repeat the above 

procedure if necessary.  Contact your local distributor or the factory if you have any questions. 

 

5. MAINTENANCE AND TROUBLESHOOTING 

No routine maintenance is required other than to keep the interior of the unit clean and free of dirt, moisture and 
other contaminants. 
The LS7000/2 consists of three main sub-assemblies.  These are the enclosure with the probe, the power supply 
board and the sensing card.  The following troubleshooting guide will assist in determining how to correct most of 
the problems, which may occur in the field

 

PROBLEM POSSIBLE 

CAUSE SOLUTION 

RED LED cannot be adjusted to 
turn on. 

Antenna lead not connected. 
No power to unit. 
Blown onboard fuse. 
Bad sensing card. 

Plug antenna lead into probe. 
Check power to unit. 
Replace fuse. 
Repair/replace card. 

RED LED remains on at all times 

Antenna probe is shorted to case 
or ground. 
 
 
 
 
Bad sensing card. 

Unplug module and position it so 
that the bare end is not touching 
anything.  Turn calibration pot 20 
turns CCW.  LED should go out.  
If so, repair, replace or clean 
antenna probe. 
Repair or replace card. 

High & low setpoints cannot be 
separated. 

Bare steel probe being used on 
conductive fluid. 

Insulate or replace probe. 

Unit triggers when material 
touches probe, but will not reset 
when material recedes from probe 
or unit and gives false alarm. 

Improper mount of probe. 
Improper calibration procedure. 
Excessive material build-up on 
probe. 
 
Probe is mounted in flow of 
material. 

Contact factory or dist. 
See section 4C. 
Perform “dirty probe calibration”: 
Recalibrate with built up material 
on probe.  See section 4C. 
Revise mounting or use time 
delay.  See sec. 4A. 

Unit will not detect material. 

Improper calibration.. 
Antenna lead not plugged into 
probe. 
Unit was calibrated with material 
touching probe. 

See calibration section 4A. 
Plug antenna lead into probe. 
 
Recalibrate without anything 
touching probe. 

Unit will not stay in calibration. 

Poor grounding or unit to vessel. 

Provide secure ground connection. 

Relay operates properly, but no 
signal at terminals. 

Burned or broken lands on power 
supply printed circuit board. 
 
 
Bad relay contacts. 

Turn off power.  Remove module 
from housing and inspect lands on 
printed circuit for damage.  
Replace as necessary. 
Replace relay or return for repair. 

PLEASE CONTACT THE FACTORY OR YOUR LOCAL DISTRIBUTOR IF YOU HAVE ANY 
QUESTIONS OR NEED ASSISTANCE. 

 
 

Summary of Contents for LS7000/2

Page 1: ...SW WI IT TC CH H O OW WN NE ER RS S M MA AN NU UA AL L I IN NS ST TA AL LL LA AT TI IO ON N C CA AL LI IB BR RA AT TI IO ON N T TR RO OU UB BL LE ES SH HO OO OT TI IN NG G W WA AR RR RA AN NT TY Y BAB...

Page 2: ...eration 2 3 Installation 2 A Inspection 2 B Physical Installation 2 C Removing Electronics 2 D Typical Applications and Wiring Diagrams 3 5 4 Calibration 6 A Setting S1 and S2 6 B Setting Sensitivity...

Page 3: ...ousing are all calibration adjustments and sensing electronics Whenever the LS7000 2 is used in electrically conductive materials a coated probe not bare steel rod must be used B Specifications Electr...

Page 4: ...around the location you have chosen to allow for installation Allow 5 turning radius to screw the probe in and clearance above for the length of the probe 3 CAUTION Always take the necessary safety p...

Page 5: ...3...

Page 6: ...4...

Page 7: ...5...

Page 8: ...of the unit The chart below will help you set S1 and S2 S1 S2 FILL EMPTY Power On Reset UP UP AUTO FILL RELAY OFF DOWN UP AUTO FILL RELAY ON DOWN DOWN AUTO EMPTY RELAY OFF UP DOWN AUTO EMPTY RELAY ON...

Page 9: ...replace card RED LED remains on at all times Antenna probe is shorted to case or ground Bad sensing card Unplug module and position it so that the bare end is not touching anything Turn calibration p...

Page 10: ...nditions or components which have been misused abused damaged by neglect or accident This warranty shall not apply to any components which have been altered or repaired without the prior written conse...

Reviews: