1. Safety instructions
<< Read carefully prior to initial operation start-up!!! >>
Electrical safety and conformity
The Pro-Line II PC must only be installed and brought into operation by qualified personnel.
The opening of the housing is also only to be carried out by qualified personnel. The Pro-
Line II PC must only be powered by a 230/240V mains supply with an attached protective
earth conductor. The mains supply must conform to DIN 0100 part 710. The computer is
also still connected to the mains supply when it is in a switched off condition. Only through
unplugging the computer will it be separated from the mains supply. The operational area
must comply with the requirements of EN/IEC 60601-1 or EN/IEC 60950-1. The unit must
only be operated using the original mains cable provided. An extension of the power sup
-
ply connecting cable is not permitted. If the original mains cable has to be replaced with
another cable or a longer cable, then only a cable validated by us may be used. With visible
damage to a power line this must be replaced immediately with a perfect mains cable that
has been validated by us. With transfer or resale of the unit this handbook must be passed
on with it.
! If these instructions are not complied with the unit loses its conformity and there is a
danger to life for users, patients and third persons.
Initial operation
When your Pro-Line II PC has just been delivered do not under any circumstances connect
it straight away! First unpack your new computer very carefully and remove the protective
foil. Then leave the PC standing in the room in which it is to be used for at least 2 hours so
that the computer can assimilate the room temperature and any possible condensation
moisture that has accrued can dissipate. If moisture has built up on the housing wipe this
off with a clean, antistatic cloth. As long as there is moisture on the housing the computer
must not be connected to the mains supply.
Purpose and mode of operation
This unit is suitable for the storage and processing of data. This unit is developed for long-
term use in intermittent operation.