Step
Procedure (PDTD-2 Test Setup - PD Measurement)
Art.Nr.
Connect all earth cables
• DUT (Device under Test)
• HVA
• between PD-2 Filter and PD-2 CC
-> should be as short as possible - if possible use for earthing
• between PD-2 CC and DUT -> should be as short as possible and at the same point
GH0522
GH0575 & GH0576
GH0576
Connect HV cables
• between HVA and PD-2 Filter -> use PD impedance
1
• between PD-2 Filter and PD-2 CC
• between PD-2 CC and DUT (connection should be as short as possible)
use KES0209 MC socket connection and KMD0081 corona shield for connecting to DUT
GH0801
3
, GH0581
GH0567
2
GH0550
2
with
KES0209&KMD0081
Connect Power Supply Plug
• HVA
• PD-2 (battery/mains); If PD is connected with power source battery is charging!
KEK0038
KEK0038
Connect Communication Ports • between PC and PD-2 CC via USB Cable
• between PC and HVA via Opto Converter
• Plug b2 Dongle at PC for b2 Suite Software
GH0586
KDG0011
GH0900
Verify Connections
• Check that all cables are attached securely.
Turn on all units
• Turn “ON” HVA main switch
• Turn key switch from HVA
to the “ON” position
• Turn “ON” PD main switch
1
except using HVA54-5,
2
for longer distance GH0550/ GH0551),
3
dependig on type of HVA
Summary of Contents for PD Series
Page 2: ...b2 electronic GmbH DHV0088 2...
Page 4: ...b2 electronic GmbH DHV0088 4...
Page 27: ...General Description b2 electronic GmbH DHV0088 27 PDTD200 2 CC Filter...
Page 29: ...General Description b2 electronic GmbH DHV0088 29...
Page 37: ...General Description b2 electronic GmbH DHV0088 37...
Page 58: ...PD Test Setup b2 electronic GmbH DHV0088 58 4 12 PDTD 2 Disconnection PD Measurement...
Page 65: ...Declaration of Conformity b2 electronic GmbH DHV0088 65...
Page 66: ...Declaration of Conformity b2 electronic GmbH DHV0088 66...