background image

6

Instructions for use

En

WF-1000

should stay in steady state room condition for at least 
30 minutes.

9.  Avoid measuring temperature in 30 minutes after exer-

cise, bathing, or returning from outdoor.

10. To protect the environment, dispose of empty batteries 

at appropriate collection sites according to national or 
local regulations.

11.  It is ill-advised to disassemble the thermometer.
12.  Please use the thermometer solely for its intended purpose.
13.  Carefully hold the device when in use to avoid dropping 

the device.

14. Allow one minute between successive measurements as 

slight variations may occur if measurements are taken over 
a short period of time. Use average temperatures instead.

15.  There are no absolute body temperature standards. Keep 

reliable records of your personal temperature to serve as 
a reference for judging a fever.

16.  Under any circumstances, the temperature taking result 

is ONLY for reference. Before taking any medical action, 
please consult your physician.

17.  It is recommended calibrate the device every 1 year.
18. Please keep this device away from pets, pests, and children.
19.  Danger of explosion if battery is incorrectly replaced. 

Replace only with the same type recommended by the 
manufacturer, discard used batteries according to the 
manufacturer’s instructions.

4.   CONTRAINDICATIONS

Contraindications: are not revealed

Summary of Contents for WF-1000

Page 1: ...Instructions for use EN WF 1000 Infrared Ear Forehead Thermometer...

Page 2: ......

Page 3: ...identification 7 6 Description of LCD display 8 7 What is a normal temperature 9 8 How to take your temperature 11 9 Cleaning instructions 14 10 Battery replacement 15 11 Troubleshooting 17 12 Applie...

Page 4: ...and international safety standards Thanks to the infrared technology this thermometer takes temperatures in seconds by measuring heat generated by the eardrum the surface skin of the forehead or other...

Page 5: ...l the notes listed below Any action against these notices may cause injury or affect the accuracy 1 Do not disassemble repair or remodel the thermometer 2 Be sure to clean the thermometer lens each ti...

Page 6: ...slight variations may occur if measurements are taken over a short period of time Use average temperatures instead 15 There are no absolute body temperature standards Keep reliable records of your pe...

Page 7: ...7 Instructions for use En WF 1000 5 PRODUCT IDENTIFICATION Forehead Cover cap Probe tip Power Scan button Battery compartment lid LCD...

Page 8: ...6 DESCRIPTION OF LCD DISPLAY Measurement display C or F Celsius Scale or Fahrenheit Scale Measurement in progress symbol Battery warning symbol Last memorized Forehead Scan Mode Ear Scan Mode Object S...

Page 9: ...eviously described an oral temperature is gen erally 0 5 C higher Rectal temperature will be 1 0 C higher Infrared forehead temperature readings are equivalent to oral temperature readings MEASURING M...

Page 10: ...operative procedures In all of these cases please consult your doctor Use of the thermometer on different persons can be inap propriate in the event of certain acute infectious diseases because of the...

Page 11: ...e only because the condition of skin is likely affected by the surroundings Measuring body temperature in the ear 1 Press the Power Scan button for 1 second to switch the thermometer on All symbols ap...

Page 12: ...n the forehead 1 Please note that the forehead temples must be free from sweat and cosmetics and that taking medication and skin irritations can distort the result when measuring temperature on the fo...

Page 13: ...play The screen displays Lo or Hi when the temperature measured is out of the measurement range See Figure 1 2 3 4 The mark Err appears if the operating temperature is NOT within 15 C 59 F and 35 C 95...

Page 14: ...temperature point the thermometer to the spot water surface as close to water as possible but do not immerse the thermometer in water or surface of an item bring the thermometer close to the object so...

Page 15: ...the screen displays a Lo sign the battery symbol on the screen blinks and beep sounds are triggered see Figure 7 It cannot work until the battery is replaced At this moment press any button to turn i...

Page 16: ...r part of the battery should be facing up You should still be able to see the sign when battery is installed 4 Slide the cover back on Unit is ready for immediate use Dispose of used batteries in acco...

Page 17: ...ged and fully seated against the opening of the ear canal Failure to properly position the probe may lead to a low temperature reading see How to take your temperature part The Probe lens is dirty Cle...

Page 18: ...ion The temperature mea sured is higher than 50 C 122 F Operate the thermometer only between the speci ed temperature ranges If nec essary clean the sensor tip In the event of a repeated error message...

Page 19: ...nt Part 2 Collateral standard Electromagnetic compatibility Requirements and tests 13 PRODUCT SPECIFICATIONS Measuring range 10 C 50 C 50 F 122 F Human Body 34 C 43 C 0 2 C 93 2 F 109 4 F 0 1 F Object...

Page 20: ...g 3 Out of operating temperature Lo HI Indication 4 Low battery power checking 5 Waterproof lens and probe cover free 14 THERMOMETER SET Thermometer set 1 Thermometer 2 Battery 1 CR2032 LI 3V 3 Instru...

Page 21: ...uidance RF emissions CISPR 11 Group 1 The WF 1000 Thermometer uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in near...

Page 22: ...ty test IEC 60601 test level Compli ance level Electromagnetic enviroment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete...

Page 23: ...occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagne...

Page 24: ...output powerof transmitter W Separation distance accordingto frequencyoftransmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38...

Page 25: ...ry protective cover and packaging When a manufacturing defect is revealed during the warranty period a faulty unit would be repaired or if repairing is impossible replaced with another one Manufacturi...

Page 26: ...the penetration of solid bodies with dimension greaterthan 12 5 mm Protected from the penetration ofvertically falling water drops Manufacturer s Name Article number Serial number Type BF equipment D...

Page 27: ......

Page 28: ...IM_WF_1000_EN_0320 www bwell swiss ch B Well Swiss AG Bahnhofstrasse 24 9443 Widnau Switzerland...

Reviews: