![B.Well WF-1000 Instructions For Use Manual Download Page 19](http://html1.mh-extra.com/html/b-well/wf-1000/wf-1000_instructions-for-use-manual_3131836019.webp)
19
Instructions for use
En
WF-1000
• ISO 80601-2-56
Medical electrical equipment –
Part 2-56: Particular requirements for basic safety
and essential performance of clinical thermometers
for body temperature measurement
• IEC/EN 60601-1
Medical electrical equipment –
Part 1: General requirement for safety
• IEC/EN 60601-1-2
Medical electrical equipment –
Part 2: Collateral standard: Electromagnetic
compatibility – Requirements and tests
13. PRODUCT SPECIFICATIONS
Measuring range
10°C ~ 50°C (50˚F ~ 122˚F)
Human Body:
34°C ~ 43°C ±0,2°C (93,2˚F ~ 109,4˚F ±0,1°F)
Object measurement: 10°C ~ 33,9°C / 43,1°C ~ 50°C ±5%
(50°F ~ 93,02°F / 109,58°F ~ 122°F) ±5%
Display resolution
0,1°C (0,05˚F)
Operating environment 15˚C ~ 35˚C (59˚F ~ 95˚F)
with relative humidity up to 95%
(non
condensing).
Storage / Transportation -25˚C ~ 55˚C (-13˚F ~ 131˚F)
environment
with relative humidity up to 95%
(non
condensing).
Power supply
1 × 3V CR2032 Size
Lithium
Battery