Page | 12
10.
Unload the cell 1 and move the weight on the top of the load cell that will be
addressed as 3. Then press calibrate / “coefficienti calcolati”.
11.
8. Unload the cell 3 and move the weight on the top of the load cell that will
be addressed as 4. Then press calibrate / “coefficienti calcolati”.
12.
9. Unload the cell 4 and move the weight on the top of the load cell that will
be addressed as 2. Then press calibrate / “coefficienti calcolati”.
13.
10. The calibration of the 4 load cells is finished. This procedure is also used
by the terminal to address the single load cells. The DD1010 ask to calculate
the Coefficients of correction. Press Calculate coefficients.
14.
The terminal shows the coefficient of correction and ask to save the values.
15.
The Eccentric calibration is finished.
16.
Follow Steps 1-7 under Analog calibration to complete a zero / span
calibration.