Floor Scale Installation Manual
23
With all of the potentiometers at “maximum signal” set the
test weight over one corner of the scale platform and record
the weight shown on the indicator. Repeat this process for the
remaining three corners. Whichever corner load cell has the
lowest reading will be used as the reference and will not be
adjusted.
Place the test weight on one of the other three corners and ad-
just the potentiometer down for that corner’s load cell to match
the weight shown for the reference cell. Repeat this for the
remaining two corners.
Adjusting the three corners may slightly change the output of
the reference cell.
Re-zero the indicator and repeat this process until all the cells / corners are /- 1% of the weight of the test weight.
Refer to your particular indicator’s manual for specific calibration procedure. In general you should place the indicator into
“calibration” mode and perform a zero calibration with no weight on the scale. Then place test weights on the scale equal
to 70%-80% of the scale’s total capacity. If several test weights are being used they should be spaced evenly around the
platform. Perform a “span” calibration on the indicator.
Remove all test weights and check the zero reading on the indicator. If necessary, repeat the above procedure for calibra
-
tion and then remove the indicator from its calibration model.
Troubleshooting
The indicator does not operate (no display):
• Check power source / plug
• Fuse in indicator has blown
• Connection cable is cut
• Incorrect connection to scale
The indicator display stays at zero:
• Faulty indicator
• Incorrect connection for load cell wires
• Load cell feet were not installed
Varying weight indication:
• Scale platform is not level
• Check for items under platform (binding)
• Load cell or cable water damage
• Vibration near scale (remove source)
• Faulty indicator
• Load foot contacting scale frame
Consistently weighing incorrectly
:
• Scale platform is binding
• Indicator not adjusted to zero properly
• Indicator not adjusted to span properly
• Load cell is damaged
• Load foot contacting scale frame
• If in a pit, debris between scale and edge