Application Notes
53
5.1.10 Statistical Mode Automatic Measurements
When operating in Statistical mode, the RFM3000 has a unique text format display that is available when the TEXT/GRAPH
system key is pressed. A sample of the text display is shown in figure
Figure 5.8
Statistical Mode Text Display
In the Statistical mode the following five automatic measurements are displayed in the RFM3000 Text display for both
input channels and both trigger channels. The Text column lists the abbreviated forms that appear on the display screen.
In the Statistical mode the following five automatic measurements are displayed in the PMX40 Text display for both
input channels and both trigger channels. The Text column lists the abbreviated forms that appear on the display screen.
TEXT
TERM
DEFINITION
Avg
Average Power The unweighted average of all linear power samples occurring since acqui-
sition started.
Peak
Peak Power
The highest power sample occurring since acquisition was started.
Min
Minimum Power The lowest power sample occurring since acquisition was started. In loga-
rithmic units a reading below the clip level will display as down arrows.
Pk2Avg
Pk/Avg Ratio
The ratio (in dB) of the Peak Power to the Average Power.
Table 5.5
Statistical Automatic Measurements