*RST
Reset the Models 885 & 886 to the power on default status. The
default status is:
1KHz 1Vrms SLOW CpD uF mH Ohm
After the Models 885 & 886 is reset, it will beep once and returns
the “BEEP” string back.
ASC
Set the format of the return value. This command sets the ASCII
string return or the numerical code.
PARAMETER:
ON
ASCII
string
OFF
Numerical
code
Example:
ASC
ON
FREQ?
1KHz
(
return value
)
ASC
OFF
FREQ?
2 (
return value
)
CORR OPEN
Perform the open calibration. This command sets the Models 885 &
886 to do the open calibration. After the calibration is done, the
Models 885 & 886 will beep once and returns the “BEEP” string
back.
CORR SHORT
Perform the short calibration. This command sets the Models 885 &
886 to do the short calibration. After the calibration is done, the
Model 885 will beep once and returns the “BEEP” string back.
34