3.2 Commands
Measurement Setting (or Querying) Command
The following measurement mode-setting and the query
commands are supported in the Model 885/886. When a
mode-setting command is entered the Model 885/886 will
return “
OK
” follows the ASCII CR (0DH) and ASCII LF
(0AH) after setting is complete. When query command is
entered, the Model 885/886 will send back the values of
measurement. After a command is entered, the meter shall
respond within 2.5 seconds with the return values follow the
ASCII CR and ASCII LF. If an illegal command is entered,
there is no response from the meter.
z
DCR(?)
DC resistance measurement mode setting or
querying command.
z
CpRp(?)
Parallel capacitance and parallel resistance
measurement mode setting or querying command.
z
CpQ(?)
Parallel capacitance and quality factor measurement
mode setting or querying command.
z
CpD(?)
Parallel capacitance and dissipation factor
measurement mode setting or querying command.
z
CsRs(?)
Serial capacitance and serial resistance
measurement mode setting or querying command.
z
CsQ(?)
Serial capacitance and quality factor measurement
mode setting or querying command.
z
CsD(?)
Serial capacitance and dissipation factor
measurement mode setting or querying command.
z
LpRp(?)
Parallel inductance and parallel resistance
32