46
10.1 Supplementary specifications – 5493C / 5493CGPIB
Display readings and measuring speed
DC resistance, DC voltage, DC current
[1]
Integration time
Digits
[2]
Readings/second
[3]
Noise error
100PLC/2s(1.67s)
6 1 2
⁄
0.5(0.6)
0
10PLC/200ms(167ms)
6 1 2
⁄
5(6)
0
1PLC/20ms(16.7ms)
5 1 2
⁄
45(55)
0.001% range
0.2PLC/4ms
5 1 2
⁄
200
0.001% range
[4]
0.02PLC/400μs
4 1 2
⁄
1000
0.01% range
[4]
[1]. Reading rate for 60 Hz and 50 Hz operating conditions, auto zero off, using a fixed range.
[2]. Digit mask set to Auto. If a fixed number of digits are selected, the number of displayed digits does
not change due to the test speed.
[3]. The average speed at which the ADC continuously samples data to the display.
[4]. Add 20 μV for DCV; add 20 mΩ for resistance. Add 0.2 uA + 10 times the above noise error range
for the 100 uA and 10 mA ranges for DC current.