3
The VenaTech® LP Brachial Introducer system (Antecubital) is intended for use
with the VenaTech® LP Vena Cava Filter.
T
c
C
During the implantation via the Brachial approach (Antecubital) the filter cartridge
must be inserted in the JUGULAR position into the cartridge cradle.
INDICATIONS
Please refer to the VenaTech® LP instructions for use booklet (JUGULAR, or
SUBCLAVIAN for OUS only, approaches) for important warnings / indications /
contraindications / implantation procedure of the VenaTech® LP Brachial
Introducer System (Antecubital). [For US market only: Vena Cava Filter Brachial
Introducer System (Antecubital)]
(1)
Note for the implantation from the Brachial approach :
•
Implantation approach:
This Introducer System is designed for insertion via the
Brachial approach
(Antecubital)
. If needed, this Introducer System can be used to insert the
filter via the right internal jugular vein or the left or right subclavian veins.
In this case, due to the length of the introducer, it is important to rigorously
follow aseptic technique.
Other approaches (particularly the left internal jugular vein, the right or
left external jugular veins) should not be attempted.
•
Preparation of the patient:
Stretch the arm of the patient 90 degrees from the long axis of
the spinal column. This will allow you to place the introducer in the axis
of the subclavian vein.
IMPLANTATION PROCEDURE
PRECAUTIONS