24
Extraoral Suction System, Instructions For Use and Installation
Statement of Manufacturer— Electromagnetic Immunity
Extraoral Suction System is designed for electromagnetic environment described below. Be sure to apply.
Electromagnetic
Immunity Test
IEC6061 Test Level
Test Level Conformance
Electromagnetic
Environment
Electrostatic
Launch (ESD)GB/T
17626.2
±
6 kV Contact Discharge
±
8 kV
Air Discharge
±
6 contact Discharge
±
8
kV Air Discharge
The floor should be of
wood, concrete, or tile. If
the floor is covered by
synthetic materials, the
relative humidity should be
at least 30%.
Electrical Fast
Transient Burst
GB/T 17626.4
±
2 kV to power wire
±
2 kV to power wire
The power supply should
reach the standard of
typical commercial or
hospital power supply.
SurgeGB/T 17626.5
±
1 kV wire to wire
±
2 kV
ground to ground
±
1 kV wire to wire
±
2 kV
ground to ground
The power supply should
reach the standard of
typical commercial or
hospital power supply.
Voltage sags, short
interruptions and
voltage changes in
power input
lineGB/T 17626. 11
< 5% U
t
,
lasting 0.5 Cycle
(At U
t
>95% Sag)40 %U
t
,
lasting
5 Cycles (At U
t,
60% Sag)70% U
t
,
lasting 25 Cycles (At U
t,
30% Sag)
< 5% U
t
,
lasting 5s (At U
t,
>95%
Sag)
< 5% U
t,
lasting 0.5 Cycle
(At U
t,
>95% Sag)40 %U
t
,
lasting 5 cycles (At U
t,
60%
Sag)70% U
t
,
lasting 25
cycles (At U
t
,
30% Sag)
< 5% U
t
,
lasting 5s
(At U
t,
>95% Sag)
The power supply should
reach the standard of
typical commercial or
hospital power supply.If
required to use the
equipment during power
blackout, it is
recommended to use
battery or uninterruptible
power supply.
Power Frequency
Magnetic Field
(50/60 Hz)GB/T
17626.8
3 A/m
3 A/m
Power frequency magnetic
field should be at the same
level with PFMF in typical
commercial or hospital
environment.
Note:U
t
refers to the AC network voltage before the test.
Summary of Contents for Optima EOS350
Page 9: ...6 Extraoral Suction System Instructions For Use and Installation 4 1 Product Dimensions inch ...
Page 10: ...7 Extraoral Suction System Instructions For Use and Installation ...
Page 14: ...Extraoral Suction System Instructions For Use and Installation ...
Page 41: ...Système d aspiration extra orale mode d emploi et installation 38 ...
Page 45: ...Système d aspiration extra orale mode d emploi et installation 42 ...
Page 72: ...Extraorales Absaugsystem Gebrauchsanweisung und Installation 69 4 1 Produktabmessungen Zoll ...
Page 73: ...Extraorales Absaugsystem Gebrauchsanweisung und Installation 70 ...
Page 77: ...Extraorales Absaugsystem Gebrauchsanweisung und Installation 74 ...
Page 106: ...Sistema de succión extraoral Instrucciones de uso e instalación 103 ...
Page 110: ...Sistema de succión extraoral Instrucciones de uso e instalación 107 ...
Page 138: ...Sistema di aspirazione extraorale Istruzioni per l uso e l installazione 135 ...
Page 142: ...Sistema di aspirazione extraorale Istruzioni per l uso e l installazione 139 ...
Page 170: ...Extraoraal afzuigsysteem Gebruiks en installatieaanwijzingen 167 ...
Page 174: ...Extraoraal afzuigsysteem Gebruiks en installatieaanwijzingen 171 ...
Page 201: ...System odsysania wydzieliny z jamy ustnej instrukcja obsługi i instalacji 198 ...
Page 205: ...System odsysania wydzieliny z jamy ustnej instrukcja obsługi i instalacji 202 ...