Communications cable specification and connection

Function list

Error indication and correction

Communications commands

Handling Precautions

• Underscores ( ) in the send examples stand

for spaces.

• The following example shows a response if an

error occurs.

NG (CR/LF)
21:illegal type

@A

Function: 

Reads out the current flow rate 
(channels 1-4).

Command example: @A (CR/LF)
Response example: 1=-3.00 (CR/LF)

2=3.00 (CR/LF)

(CR/LF)

This shows that no sensor is
connected to ch. 3.

4=0.00 (CR/LF)
(CR/LF)

@PRE

Function: 

Sets the ON/OFF point for any channel.

Command example: @PRE11

-3.00 (CR/LF)

Response example: OK (CR/LF)

@PRE11: Channel 1, L1/

L setting

@PRE12: Channel 1, L2 setting
@PRE21: Channel 2, L1/

L setting

@PRE22: Channel 2, L2 setting
@PRE31: Channel 3, L1/

L setting

@PRE32: Channel 3, L2 setting
@PRE41: Channel 4, L1/

L setting

@PRE42: Channel 4, L2 setting

@HYS

Function: 

Sets the hysteresis width for any
channel.

Command example: @HYS11

0.02 (CR/LF)

Response example: OK (CR/LF)

@HYS11: Channel 1, hysteresis width setting
@HYS21: Channel 2, hysteresis width setting
@HYS31: Channel 3, hysteresis width setting
@HYS41: Channel 4, hysteresis width setting

@P

Function: 

Sets trigger signal for memorizing
reference flow value is enabled or not,
when window comparator mode 2 or 3
is selected as the operation mode.

Command example: @P1 (CR/LF)
Response example: OK (CR/LF)

@P1: Trigger signal for memorizing reference flow value

is enabled at channel 1

@P2: Trigger signal for memorizing reference flow value

is enabled at channel 2

@P3: Trigger signal for memorizing reference flow value

is enabled at channel 3

@P4: Trigger signal for memorizing reference flow value

is enabled at channel 4

@MODE

Function: 

Sets the operation mode for any
channel.

Command example: @MODE1

1 (CR/LF)

Response example: OK (CR/LF)

@TYPE

Function: 

Sets the flow sensor type for any
channel.

Command example: @TYPE1

1 (CR/LF)

Response example: OK (CR/LF)

@TYPE1    1

1: Ch. 1
2: Ch. 2
3: Ch. 3
4: Ch. 4

1: -3.00 to +3.00L/min
2: 0.00 to 3.00L/min
3: -500 to +500mL/min
4: 0 to 500mL/min
5: 0.00 to 10.00L/min

]

@MODE1   1

1: Ch. 1
2: Ch. 2
3: Ch. 3
4: Ch. 4

1: Window comparator mode 1
2: Window comparator mode 2
3: Window comparator mode 3
4: Window comparator mode 4

]

@P1

1: Ch. 1
2: Ch. 2
3: Ch. 3
4: Ch. 4

@HYS1 1

1: Ch. 1
2: Ch. 2
3: Ch. 3
4: Ch. 4

1  fixed

]

@PRE1 1   

1: Ch. 1
2: Ch. 2
3: Ch. 3
4: Ch. 4

1: L1/

L

2: L2

]

]

Error details

Error indication

Correction method

(“n” shows relevant channel.)

(“n” shows relevant channel.)

OFF

E-

 

1

E-2

n

n

E-3

Flow sensor of the selected 
channel is not connected or 
wiring is faulty.

Replace the flow sensor   
after turn OFF the power 
or the cable. 

After correcting the  
error, keep the mode key    

pressed for 1 second 

or longer.

Threshold flow value 
beyond the measurement 
range is set in window 
comparator mode 2 or 3.
Excessive voltage (5V or 
more) is input to the sensor 
(channels 1-4).
Output current is excessive.

Signal name

DC24V

GND
SW1
SW2
SW3
SW4

SG

RXD

TXD

N.C
N.C

Pin No.

1
2
3
4
5
6
7
8
9

10
11

Signal name

D.G

DCD
RXD

TXD
DTR

DSR

RTS
CTS

Pin No.

5
1
2
3
4
6
7
8

This device

Personal computer side

D-sub 9-pin configuration

Serial communications
cable

Power and output cable

Signal name

DC24V

GND
SW1
SW2
SW3
SW4

SG

RXD

TXD

N.C
N.C

This device

Personal computer side

D-sub 25-pin configuration

Pin No.

1
2
3
4
5
6
7
8
9

10
11

Signal name

D.G

TXD

RXD

RTS
CTS

DSR
DCD
DTR

Pin No.

7
2
3
4
5
6
8

20

Serial communications
cable

Power and output cable

Function

Key

RS-232C

operation

command

Instantaneous flow rate indication

Yes

@A

L1 (

L) / (L2) settings

Yes

@PRE

Hysteresis settings

Yes

@HYS

Trigger signal for memorizing 

Yes

@P

reference flow value
Operation mode selection

Yes

@MODE

Flow sensor selection

Yes

@TYPE

Zero reset setting

Yes

@B

Auto-scanning indication

Yes

@AS

Output mode setting

Yes

@INV

(only for comparator mode 4)
Peak hold setting

Yes

@PHL

Bottom hold setting

Yes

@BHL

Indication on/off

Yes

@DIS

Operation mode confirmation

No

@MD

Flow sensor type confirmation

No

@TP

L1 / L2 confirmation

No

@C

Threshold flow value confirmation

No

@E

(only for modes 2 and 3)

Output status confirmation

No

@SW

Switch output response time setting

Yes

@DLY

Switch output response time confirmation

No

@SD

Flow rate display moving average setting

Yes

@FIL

Flow rate display moving average confirmation

No

@F

Summary of Contents for MCW400A100

Page 1: ...thout notice Considerable effort has been made to ensure that this manual is free from inaccuracies and omissions If you should find an error or omission please contact the azbil Group In no event is...

Page 2: ...ion part stands above the connector main body as shown in the figure below If the connector cover is not lifted off the connector main body it cannot be used 1 Cut the cable to the required length Rem...

Page 3: ...scanning indication Auto scanning indication at all channels is available Handling Precautions Since the auto scanning mode is also used for the key lock function key operations other than auto scann...

Page 4: ...h channel Hysteresis settings H S 1 Ch 1 H S3 Ch 3 H S2 Ch 2 H S4 Ch 4 Handling Precautions To prevent chattering set the hysteresis value to 2 or more 3 Advanced settings 3 1 Reading the reference fl...

Page 5: ...erase write cycles to exceed the guaranteed amount Window comparator mode 4 L1 and L2 can be set to desired levels Hardware requirement PC PC 98 series or equivalent DOS V machine OS Windows 95 or lat...

Page 6: ...0L min MODE1 1 1 Ch 1 2 Ch 2 3 Ch 3 4 Ch 4 1 Window comparator mode 1 2 Window comparator mode 2 3 Window comparator mode 3 4 Window comparator mode 4 P1 1 Ch 1 2 Ch 2 3 Ch 3 4 Ch 4 HYS1 1 1 Ch 1 2 Ch...

Page 7: ...on ON or OFF Command example BHL 1 CR LF Response example OK CR LF DLY Function Sets the switch output response time for any channel Command example DLY1 2 CR LF Response example OK CR LF SD Function...

Page 8: ...without condensation without freezing Storage temperature 20 to 80 C without condensation without freezing Dielectric strength 500Vac 1min Insulation resistance 100M or more 500Vdc megger Vibration re...

Reviews: