6-18
Chapter 6. Communication Data
Item
Data range
/ [unit]
Data address
Decimal
(hexadecimal)
R
W
Notes
(P-29)
Automatic display off
time
0 to 9999 [s]
2229 (08B5)
✓
✓
When set to “0,” the display is always
lit. (display turnoff function is
disabled)
(P-30)
PV display filter time
constant
0 to 9999 [ms]
2230 (08B6)
✓
✓
First-order lag low-pass filter for
display output
(P-31)
Keylock password
0 to 9999
2231 (08B7)
✓
✓
(P-32)
Analog scaling 2
(10 to 100 % FS)*
2
2232 (08B8)
✓
✓
The value is without a decimal
point.*
1
(P-33)
Totalization pulse
weight
[1/pulse]
2233 (08B9)
✓
✓
The value is without a decimal
point.*
1
(P-34)
Totalization pulse width
20 to 100 [ms]
2234 (08BA)
✓
✓
Set in multiples of 10.
*1. For the number of decimal places and the unit of data, see “Flow rate digits after the decimal point” (address 1003) and
“Flow rate unit” (address 1005).
Ex.: If the flow rate digits after the decimal point is set to “1” and the flow rate unit is set to “0” (mL/min), the value “1234”
means 123.4 mL/min.
*2. The setting range is the full scale flow rate (address 1002) multiplied by the percentage in parentheses. (The setting
range varies depending on the model and gas type.)
*3. If the data exceeds the range, the maximum value of the range is read.
*4. For the number of decimal places and the unit of data, see “Totalized flow rate digits after the decimal point” (address
1004) and “Totalized flow rate unit” (address 1006).