7.

 

  Preparing working environment 

 

Location for the scale should be chosen with care in order to limit influence of the factors that can interrupt 
working scale. This location has to maintain proper temperature for working scale and necessary space for 
its operating. The scale should stay on stable table made of material that does not influence magnetically 
on the scale. 

Rapid air blows, vibrations, dust, rapid temperature changes or air humidity over 90% are not allowed in 
scale surrounding. The scale should be far from heat sources and devices emitting strong electromagnetic 
or magnetic fields. 

 

8.

 

Preparing scale to work 

 

 

1.

 

Take the scale out of the package removing protective 
foils. It is recommended to keep the original package in 
order to transport the balance safely in future. 

 

2.

 

Take  a  pan  off  and  remove  protective  elements  from 
below the pan. 

 

3.

 

Place  the  scale  on  a  stable  ground  not  affected  by 

mechanical vibrations and airflows. 

4.

 

 Level the scale with the rotating rear legs so that the 
air bubble in the water-level at the back of the scale is 
in the middle.  

  

 

 

  Correct 

Wrong 

Summary of Contents for BD2P

Page 1: ...USER MANUAL BD2P balance File 2016 04 19 BD_P bW109 GB AXIS 80 125 Gdańsk ul Kartuska 375B POLAND www axis pl ...

Page 2: ...ht unit selection UnIt 27 18 4 Serial port parameters setting SErIAL 28 18 5 Printout configuration PrInt 29 18 6 Setting backlight function b_LIGHt 31 18 7 Analog out configuration AnALoG 32 18 8 Entering reference zero value ZErO 33 19 Special functions description 34 19 1 Products and users database Prod i USEr 35 19 2 Pieces counting function PCS 39 19 3 Percentage weighing function PErC 40 19...

Page 3: ...casing against opening notified body stamps and green metrological marking placed on the balance name plate Renewing of scale legal verification is required when protective seal is violated or after period of 3 years starting from 1st December of year when first legal verification was performed In order to renew legal verification please contact authorized service of AXIS NACE classification 29 24...

Page 4: ...____________________________________ 3 Technical data Scale type BD2P Max load 2kg Min load 5g Readout d 1g Verification plot e 1g Accuracy class III Working temperature 10 40 C Tare range 2kg Weighing time 3s Pan size 225x165mm Scale size 235x250x65mm Supply 230V 50Hz 6VA 12V 0 16A Scale weight 3 5kg ...

Page 5: ...l functions menu key print out key HR increased indication resolution indicator 0 zero indicator indicator result stabilisation indicator indicator NET net weight indicator indication with subtracted tare indicator MODE special function setting bar indicator total load indicator graduated 0 100 indicator OFF standby indicator B G gross mass key B G indicator indicator Pcs lb t oz kg pieces countin...

Page 6: ...te type supplied with the scale Pay attention that supply voltage is compatible with specified technical data Do not use the scale when its cover is opened Do not use the scale in explosive conditions Do not use the scale in high humidity If the scale seems not to operate properly unplug it from the mains and do not use until checked by authorised service 6 Environment protection According to lega...

Page 7: ...id air blows vibrations dust rapid temperature changes or air humidity over 90 are not allowed in scale surrounding The scale should be far from heat sources and devices emitting strong electromagnetic or magnetic fields 8 Preparing scale to work 1 Take the scale out of the package removing protective foils It is recommended to keep the original package in order to transport the balance safely in ...

Page 8: ...placed on the pan To make the control of a load on the pan easier and to avoid exceeding measurement range the scales have load indicator calibrated 0 100 4 Weighing result should be read when the indicator lights which signalises result stabilisation 5 When the scale is not used but it is necessary for it to be ready to work it can be switched off by pressing I key The scale reading system is the...

Page 9: ... Auto OFF functions 3 Charging accumulators is performed automatically after connecting feeder to the scale also during weighing Accumulator power level can be read using bAttErY function bat VoL option 11 Scale checking It is advised to check scale indication accuracy before and after series of measurement using any load with known weight To check the scale with legal verification use a calibrati...

Page 10: ...urements Office Performing calibration requires changing adjustment switch position which is placed behind protecting mark sticker of a notified body An access to the switch is possible only after removing the mark Therefore scale calibration causes lost of legal verification and in consequence the necessity of renew legal verification in the nearest notified body or in place where the balance is ...

Page 11: ...lock is installed and on and nett weight During transmission LabEL communicate is displayed The way of sending data and transmission parameters is set using SErIAL special function Set of send data is set using special function PrInt The following data can be send Header scale type Max d e serial number Operator identification number Successive printout number measurement Identification number or ...

Page 12: ...itiaton signal Scale Computer scale send indicator S stable or U unstable 16 bytes the same as SI command Attention Network number different than zero SErIAL nr function changes scale working mode communication with a computer is possible after logging the scale in with 02h scale number command To log the scale out use 03h command For example Using a program to test RS232 interface program is avai...

Page 13: ...digits after dot should be the same as on scale display Scale Computer without response Example in order to set low threshold 1000g in scale B1 5 d 0 5g the following order should be sent S L 1 0 0 0 0 CR LF 53h 4Ch 31h 30h 30h 30h 2Eh 30h 0Dh 0Ah in order to set low threshold 100kg in scale B150 d 50g the following order should be sent S L 1 0 0 0 0 CR LF 53h 4Ch 31h 30h 30h 2Eh 30h 30h 0Dh 0Ah S...

Page 14: ...variable signs mm gg 5 signs minutes hour rrrr mm dd 10 signs year month day masa 10 signs scale indication mass unit P1 Steering instruction Attention 1 Except variable signs constant signs can also be inscribed e g factory name product name and so on 2 In standard only one label pattern is possible to printout number 0001 Using bigger amount of patterns other label numbers is possible thanks to ...

Page 15: ...pan empty plug the scale to the mains with a ground contact The scale proceeds with following start up actions Display test Meter type displaying Program version The scale is ready to work Attention UnLOAd communicate means that the scale is loaded or transportation security elements situated under the pan has not been removed 0 0 kg kg Zasilacz 230V 12V ...

Page 16: ... Zero indication and 0 indicator mean that the scale is ready to work After putting container package tare the scale using T key NET indicator will show up Put on weighted object and readout net weight NET indicator shows that scale indicates net weight In order to readout gross weight press key B G indicator shows that scale indicates gross weight Press again key in order to come back to net indi...

Page 17: ...out of the box scale after pressing MENU key only SEtuP option it contains all configuration options is available One of the configuration options is Menu that is used to create personalized menu To add a function to personalized menu press T key when the function is indicating Chosen function is indicated with o sign on the left side of display After adding all necessary functions press out in or...

Page 18: ...ds and then the scale sequentially displays next menu positions Choosing menu position option is done by pressing T key when it is displayed on the screen After choosing position option usually several options show up on turning on selected option OFF turning off out out to menu Accelerated working with menu Menu first position is displayed for about 7s During this time user can view next position...

Page 19: ...orking method During standard weighing key is used to switch between net and gross indication When special function e g PCS is turned on using key enables to go back to standard weighing mode Sign o on the left side signalizes that special function is turned on and user can go back to function mode by pressing key 0 0 0 kg 0 0 0 kg kg kg ...

Page 20: ...normal weighing temporary 5s readout resolution increase is possible Return to normal indication is made automatically Inscribing numerical values Inscribing numerical values is needed in some special functions e g tArE function requires to inscribe tare values Keys 0 increasing digit value decimal point T next digit position MENU end of inscribing 0 0 kg kg 5s 0 kg ...

Page 21: ...on exit autozeroing autozeroing on autozeroing off lasts 10min exit settings creating menu added to menu choosing product choosing user pieces counting function actual unit selection percentage conversion function animals weighing function tare memory bank maximum value function force and torque value function summation series of measurements function threshold values comparing function statistics...

Page 22: ...nsmission type Stab no StAb Auto Cont rEMoVE protocol LonG EPL EPL_A EPL d PEn 01 cooperation with barcode readers MJ 4209 port 2 as above printout data configuration transmission port 1 header scale type Max d e serial nr scale operator id number user name succesive printout measurement number product identification number product barcode number product name pieces quantity PCS function product u...

Page 23: ...________________________ bAttEry AUto OFF ZEro dEFAULt SErVICE out turn on off accumulator charging if the scale is equipped with accumulator automatic turning off saving accumulator power as above scale start zero inscribing factory zero restore default settings for all options options only for service exit ...

Page 24: ... user menu CALIb scale sensitivity calibration AutoZEro ing self maintaining zero indication unloaded scale UnIt weight unit selection SErIAL setting serial ports Print transmission printout data selection FILtEr anti disturbance filter b_LIGHt backlight setting Ad420 analogue out configuration FIrMW are updating software only for service dEFAULt reset to factory settings sample of using in chapte...

Page 25: ... StP calibration with external weight confirmation of succesive steps MENU key out leave without changes Press T key when CAL StP option appears calibration in two steps Press T key when weight value used for calibration is indicating or use othEr option and inscribe proper value keys 0 T Press MENU and wait for writing zero to the scale When LOAD message appears put standard of mass on the pan Pr...

Page 26: ...ble zero indication if the pan is empty or if zero indication was acquired by pressing T key To turn on the function use MENU key and using T key choose AutoZEr and then Aut on To leave the function press MENU key then with T key chose AutoZEr and Aut OFF Note 1 AUt sign occurs only in scales with LCD display 2 In scales with 0 key active function changes name into AutoZE autozeroing and works onl...

Page 27: ...0 001g milligram KGrAM 1kg 1000g kilogram Pound 1 lb 453 592374g English pound OunCE 1oz 28 349523g ounce OunCEt 1ozt 31 1034763g pharmaceutical ounce GrAIn 1gr 0 06479891g grain PennYW 1dwt 1 55517384g jewellery mass unit GrAM 1g gram The way of choosing carats as weighing unit is shown on the example Attention In scale with LED display designations of mass units lb kg oz ozt ct are not displayed...

Page 28: ...ar bitS 7 8 parity control PArItY nonE no control Odd nonparity Even parity control scale number in network nr if the scale doesn t work in network the number must be 0 transmission through serial interface SendInG StAb transmission after key is used and result is stable noStAb transmission after key is pressed without need of stabilisation Auto automatic transmission after load is put on and resu...

Page 29: ...t PCS function APW unitary mass PCS function netto net mass tArE current tare value GroSS gross mass totAL total mass totAL function In printout configuration user can set if measurement printout number is saved after turning off the scale or not Enter option Print and choose Prn no Following options will appear rESEt resetting zeroing measurement number counter SAVE activate saving measurement nu...

Page 30: ...ntout during normal weighing with clock option all printout positions deactivated Sample printout during normal weighing some printout positions activated 20 07 kg 20 04 kg 20 04 kg 20 07 kg 2012 11 08 10 01 20 04 kg 2012 11 08 10 01 20 04 kg 2012 11 08 10 01 BA30 MAX 30kg e d 0 01kg S N ID OPER 000001 DATE 2012 11 08 TIME 12 26 NO 3 ID PROD 01 COUNT 0 PCS APW 0 000 g NET 3 08 kg TARE 0 00 kg GROS...

Page 31: ...he work mode of scale display backlight b_L OFF switch backlight off b_L on switch backlight permanently on b_L ECO switch off after 30 seconds of inactivity no load changes and no key operation b_L bAt like above but when powering from accumulators only out out without changes Switching backlight off causes decrease of energy consumption by the scale what is important during powering from accumul...

Page 32: ..._________________________ 18 7 Analog out configuration AnALoG This option enables to set up analog out 4 20mA or 0 10V working method used e g in PLC regulators AnG rnG inscribing Max value AnG CFG working mode configuration PLUS workmode for only positive values MinuS only for negative values both for both 0 kg ...

Page 33: ... MENU key When ZErO is displayed press T key On the display a sign ZEr Cod will show up momentary and the a dash on last digit position To enter code in new scale 1234 use keys 0 increasing digit T next digit MENU end of inscribing The following options appear successively on display ZEr Cod enter new secure code value ZEr SEt enter new zero value Using T key choose ZEr SEt Direct result from A C ...

Page 34: ...n tArE maximum value indication function UP force measuring function nEWton statistical calculations StAt paperweight calculation function PAPEr and functions that require additional equipment to be completely functional option with accumulator supply Setting accumulators charging bAttErY Automatic switching off scale function AutoOFF options with the clock setting current date and time function d...

Page 35: ...asheet form where each product has one row and each column have product data This way created database saved in csv extension with semicolons can be send to scale using Scale Database software and scale s serial interface Scale Database is available on our webpage www axis pl en Database and possibility to cooperate with external devices printer label printer barcode reader and computer enables to...

Page 36: ...abase then no communicate appears the scale stores id barcode number temporarily in memory and enables to send it to the port to printer computer together with actual weighing result LIMIt searching through products from database if product is found then Found communicate appears and all product data is recalled if there is no product in database then not Found communicate appears Prn_P sending al...

Page 37: ...from outside the base accepted Std mode not Found barcode from outside the base not accepted and no product is selected LIMIt mode Found product barcode found in database and data recalled Attention If the scale doesn t indicate any communicate check barcode reader connections port configuration and transmission protocol SErIAL function Other fast way is to press and hold MENU key about 3s Prod Id...

Page 38: ...ing product is equivalent with choosing corresponding label number If database is not used proper label can be choosed using LabeL option To each weighing results transmission a set of product and user identification data is added The set is activated in Print option Available data from products and users base Print SEtuP option USEr Id user identification number USEr nA user name inscribed from P...

Page 39: ...S APW set unitary mass directly PCS rS inserting number of details in a sample and receiving of their mass from other scale connected by RS 232C It is advised that single piece weight is not less than one reading unit and sample weight used in first phase is bigger than 100 reading units To leave function press MENU key and then using T key chose PCS and PCS oFF Note 1 APW too LOW communicate sign...

Page 40: ...100 second phase measuring specific sample as a percentage of the reference sample Weighing result is displayed in different format depending on the weight value of reference sample The function has the following options PEr oFF disable the function PEr on set current scale indication as 100 and activate percentage weighing out exit without changing settings Note 1 PEr Err message informs that ref...

Page 41: ...r should be saved in scale memory according to printer manual Label pattern choice is made by inscribing label number using LAbEL function Press MENU button When LAbEL is displayed press T key Actual label number will show To enter new label number press T key to exit function without number change press MENU To inscribe label number use keys 0 digit increase T next digit MENU end After entering l...

Page 42: ...ding the scale LOC Prn the measurement initiated manually by pressing key When LOC on is displayed press T key Tare the scale using T key if necessary and place the animal on the pan Wait until the weighing result is averaged scale display blinks Then scale will show stable averaged result and will send it through serial port The result remains on display for about 30 second Important notes 1 The ...

Page 43: ...function allows holding maximum or minimum value that is indicating at the moment Before measurement scale should be tared Function has following options UP oFF function off HIGH holding maximum value LOW holding minimum value Pressing T key will cause result zeroing Note Autozeroing function and the stabilisation indicator are deactivated when UP function is running 0 0 g kg 1kg ...

Page 44: ...vation causes displaying result in force units N Press MENU key Using T key choose NEWto function Function has several options nEW oFF function off nEW on measurement in Newtons ArM torque measurement arm length should be inscribed in meters using T 0 and MENU keys Attention Units convertion from mass kg to force N is made for acceleration of gravity g 9 80665m s2 Note 1N 0 1019kg 0 g 0 ...

Page 45: ...ot working with receipt printout after each measurement tot working without receipt printout tot CFG saving measurement mode using key Manual after taking off the load auto Press T key when tot is displayed Perform measurement series by pressing key for storing results into total register In order to print and display results enter the function by choosing totAL and tot Prn option from menu The re...

Page 46: ...ceipt form Report form total weight number of measurements average weight Note When the scale doesn t have an internal clock Date and Time do not appear on printout Maximum number of measurements is 99 999 Maximum total load 99 999 000d The weighing unit of the total value from the register Total is the same as the weighing unit stated on the keypad or is 1000 times greater what is signalled by o ...

Page 47: ...l greater than upper threshold the scale signals MAX red colour long sound signal The checkweighing results can be use to control optical indicator Indication mode batching devices Batching mode Standard scale is set for cooperation with optical indicator On outputs P1 P3 Relays socket short circuit states appear as result of comparison scale indication with threshold values On the chart below out...

Page 48: ... T key The following options for entering thresholds are displayed SEt 0 go to weighing with signalling threshold excess SEt 1 set lower threshold value SEt 2 set upper threshold value SEt 3 set zero signalisation threshold Using T key select SEt 1 option Set lower threshold value using the following keys 0 digit increase decimal point T move to next digit MENU finish Then select SEt 2 option and ...

Page 49: ... e g 1N4148 It is advised to use MS3K P electronic board sold separately consisting of RM96P transmitters with DC24V input voltage and AC250V 3A output Important notes 1 After switching the scale on both thresholds are set to maximum values 2 When setting upper threshold value pay attention that its value is not below lower threshold value 3 Setting lower and upper threshold value is possible afte...

Page 50: ...At SEt change current date and time dAt PIn data and time secure password to prevent from changing date and time by unauthorized personel dAt For data printout in different format The example at the left presents how to set current date and time using dAt SEt option After setting proper date and time activate it with dAt on option Date and time format PL rrrr mm dd gg mm UE dd mm rrrr gg mm USA mm...

Page 51: ...cale and a pilot In scale and in pilot the same radio channels must be chosen Function should be used when communication is disturbed by other devices that use the same communication channel Press MENU key and choose rF CHn by pressing T key The following communicates will appear on display Na wyświetlaczu pojawią się kolejno CHn 01 channel 1 CHn 02 channel 2 CHn 16 channel 16 out out without chan...

Page 52: ...during work with feeder and checking their power level The function has the following options bAt OFF charging off option required if ordinary batteries are used bAt on charging on accumulators are being charged even after switching scale off using I key bAt VoL reading power level of accumulators in go back to mass indication pressing MENU key out leave without changes An attempt of charging ordi...

Page 53: ...helpful in scales supplied from accumulators The function causes scale to switch off automatically Press MENU key When AutoOFF is displayed press T key The following options appear successively on display AOF oFF deactivate function AOF on activate function scale turns off after 5 minutes of not making any actions AOF bAt as above but only when supplied from accumulators Out out without changes 0 ...

Page 54: ...ur after the scale is loaded and its indications stabilize After each loading printout is made with number of measurements result date and time if clock is installed and the function is activated For the obtained measurements series the scale evaluates n number of samples sum x sum of all samples n x x sum _ x average value sum x n min minimal value from n samples max maximal value from n samples ...

Page 55: ...ng tolerance in Sta tAr automatic tare on off StA CFG function configuration Auto Automatic work samples are confirmed after loading the scale and indication stabilization ManuAL manual work confirmation is made by pressing key out exit from function Remember first to inscribe nominal weight value and tolerance mentioned above After that push T key when StA o is displayed Put on successive objects...

Page 56: ...RAGE average weight as Total n MIN minimum weight in n samples MAX maximum weight in n samples ST DEV standard deviation ST DEV standard deviation percentage To finish work with this function and zeroing result register press MENU key and then when StAt and Sta oFF is displayed press T button Statistics function cooperation with computer and Printer Scale can be equipped with two serial ports mark...

Page 57: ...nu To enter the function press T key when PAPEr is displayed Following options show on the display PAP oFF turn off the function PAP on turn on PAP n inscribing number of paper pieces on pan PAP ArE inscribing surface of single piece in m2 Press T key when PAP n and PAP ArE is displayed Enter number of samples using 0 increasing digit T next digit MENU end of inscribing Press T key when PAP ArE is...

Page 58: ...nt for repair as messenger delivery only in original package if not there is a risk of damaging the balance and loosing guarantee Error communicates Communicate Possible cause Remedy C 1 6 over 1 min selftests failed if displayed more than 1 minute contact an authorised service the scale was switched on with loaded pan remove a load from the pan mechanical damage of the load cell contact an author...

Page 59: ...lowing markings on the name plate 16 the number of the Notified Body responsible for EC verification 1443 two digit number of the year of EC verification a green metrology sticker with M mark a protective seal affixed by the Notified Body comply with the requirements on the Type Approval Certificate WE No T7950R0 and are verified to comply with EN 45501 2015 Additional information Conformity evalu...

Reviews: