01
Channel
2 channels with adjustable intensity
02
Intensity
Adjustabl 0 - 80 mA with a load of 500 Ohm
on each channel
03
Pulse amplitude
Asymmetrical, bi-phasic square pulse
04
Output Voltage
0 - 50 V
05
Power Source
1 pc. 9,0V battery
06
Size
10,2 cm (L) x 6,2 cm (W) x 3,4 cm (H)
07
Weight
150 g incl. battery
08
Frequency
Adjustable from 2 - 150Hz
09
Pulse width
Adjustable from 50 - 300μs
10
Modes
5 TENS modes with adjustable parameters:
Burst, Normal, Modulation, SD1, SD2.
P-Mode with 12 pre-defined programs (P)
11
TENS
Burst mode
(B)
Burst frequency adjustable from 0.5 - 5Hz
Pulse width adjustable from 50 - 300μs
12
TENS
Normal mode
(N)
Frequency adjustable from 1 - 150Hz
Pulse width adjustable from 50 - 300μs
13
TENS
Modulation
mode
(M)
Die Frequenz wird innerhalb von 5 Sekunden
auf 50% des Einstellwertes gebracht. In den
folgenden 5 Sekunden wird der Wert wieder
auf den Einstellwert erhöht.
14
SD1 mode
(SD1)
The SD1 (Strength-Duration) mode consists of
automatic modulation inten-sity and pulse width
in 40% range. The intensity is always increasing
while the pulse width is decreasing and vice ver-
sa. The intensity is decreased by 40% while the
pulse width is increased by 40% in 5 seconds. In
the next 5 seconds, the intensity is increased by
40% while the pulse width is decreased by 40%.
Total cycle time is 10 seconds.
15
SD2 mode
(SD2)
As SD1 mode, but with 70% modulation.
16
Program mode
(P)
Explanation of the programs on p. 22 - 23
17
Low battery
indicator
A low battery indicator will show up on the
LCD when the battery is low. Replace battery
soon.
TECHNICAL DESCRIPTION
14
AXION
EN