equipment ,and even bring the risk of electronic shock. Be sure to cut off power and check input terminal and the 
position of range knob.After confirmation, that there isn’t any error, turn on power .If the current range under 
test is unknown beforehand , set the knob to the highest range ,select a proper range according to the displaying 
value. 

mA input jack ,too large current will blow the fuse ,should replace the fuse . 

The dimension of fuse is 

Φ

5×20mm,electric dimension : 200mA250v ; 

Have no fuse internal for10A input jack . 

In order to operate safely, measurement time should be within 10sec.and next measurement should be after 15 
minutes . 

4.RESISTANCE MEASUREMENT (FIG.4) 

4-1.Apply the red test lead to “V

mA” terminal and the black 

one to “COM” terminal. 
4-2.Set the fuction range knob to resistance range, multipled 
the test leads to the resistance under tested .  
4-3. Read the measured value from the LCD .  

 

NOTE:To avoid the damage of the instrument, before 
measurement ensure that power is turned off and capacitors are 
released completely. 

At 200

 range,it should make the test leads short to measure 

the wire resistance , then substract from the real measurement. 
When the measured resistance is over 1M

,the instrument 

needs a few seconds to make reading stable ,it’s normal . 

5.DIODE AND CONTINUITY TEST (FIG.5) 

5-1.Apply the red test lead to “V

mA” terminal , the black one to “COM” terminal. 

5-2.Set the knob “diode” range ,connect the red test lead to diode positive polarity,the black one to cathod 
polarity. 
5-3.Read the measured value from the LCD. 
5-4.Connect the test leads to both ends of the wire under tested .Built-in buzzer sounds when resistance between 
the two ends is lower than approx.70 

.

  

 
 
 

Summary of Contents for AX-101B

Page 1: ...DIGITAL MULTIMETER AX 101B OPERATION MANUAL...

Page 2: ...e weakened or lost 1 Before operation ensure that the insulation layer of test leads remains unwounded Do not operate the meter if test leads cable or housing insulation is obviously damaged 2 When op...

Page 3: ...function left bottom of LCD displays H Low battery LCD displays symbol Built in battery 9V NEDA1604 6F22 006P Operation temperature 0 40 32 104 Storage temperature 10 50 14 122 Dimensions 150mm 73 5mm...

Page 4: ...est range then select a proper range according to the displaying value If LCD displays 1 it means over range set the knob to a higher range On each range the input impedance of instrument is 10M such...

Page 5: ...rminal 4 2 Set the fuction range knob to resistance range multipled the test leads to the resistance under tested 4 3 Read the measured value from the LCD NOTE To avoid the damage of the instrument be...

Page 6: ...is 6 SQUARE WAVE SIGNAL OUTPUT Set the knob to square wave range the instrument output square wave between V mA and COM NOTE Square wave experiment signal with its rich harmonics can be used as a sim...

Page 7: ...protection 500 V DC or AC peak value Frequency response 40 400 Hz Display sine wave RMS 3 DCA RANGE ACCURACY RESOLUTION 2mA 1 0 2 1 A 20mA 10 A 200mA 1 2 2 100 A 10A 2 0 5 10mA Max input current 10A l...

Page 8: ...m circuit under tested and input jack Set the knob to off gears and turn off the power 2 Loosen the screws on battery case by screwdriver and remove the battery case 3 Take out the old battery and rep...

Reviews: