4. Operating Instructions
Press the measuring trigger to power on, press again for IRT scan,release it to hold data. Press the MODE button to switch
among AT, DP and RH. Hold the trigger then press MODE to adjust the emissivity 0.75, 0.85, 0.95; Long time holding
the MODE to power off. Press once to control the laser to power on and off. Press unit switch button to change the unit.
Hold this button to change to mute condition The green LED, means low possibility of mildew; Orange indicates there
will be mildew possibility. Red LED means the high possibility of mildew. In certain humidity condition, if the target
temperature is close or below to the dew point temperature, the possibility of of mildew is greater.
5. Battery Installation
Open the battery compartment, insert one 9V battery (6F22 or 6LR61), close the lid.
6. Temperature measurement
Point the product to target objects, hold the trigger to read the temperature. The distance and target area: the value of D
(distance): S (spot) must be considered (as showed), the ratio should be no more than the value of D:S. The field of view
must be full of objects. As the distance from the object increases, the spot size of measuring area becomes larger.
Field of view: Be sure the target area is larger than the unit’s spot size. The smaller the target get, the closer the measured
distance. For precise measuring, make sure the target is at least twice as large as the spot size. The light-emitting device