KG-Series Manual

4. APPENDIX

4.1.  RS232 Serial Data Communication

This balance has a Bi-Directional RS232 port for communication with a PC. When connected to a computer, 

the weight value can be sent to the computer and commands can be sent to the balance

Transfer Format: Asynchronous

Communication Protocol: 1 start bit, 8 data bits, 1 stop bit

Baud Rate: 9600

Parity: None

Flow Control: Xon/Xoff 

When sending the weight request command to the scale, the scale’s serial port will reply with the current 

weight in ASCII format with the following structure:

1

2

3

4

5

6

7

8

9

10

11

12

13

Weight Data

Unit

Example String:  “  0 0 0 1 2 3 . 4 5 c t  ”

Hex Output: 11h 20h 30h 30h 30h 31h 32h 33h 34h 35h 63h 74h 20h 20h 13h

Meaning: 123.45ct (carats)

When sending a command to the scale such as Tare, the scale’s serial port will reply with <ACK>

NOTE:

 The CD included with the balance is meant for demonstration purposes only.  To install, simply copy 

the program files to your computer and launch Weight Display.exe. By default, the program listens for 

weight data on serial port1.

4.2. Serial Number

When communicating with the balance from a PC, a serial number (from 1-100) is used to identify the 

balance. By default, the serial number is 001. To change the serial number, press and hold the UNIT and 

MODE buttons together until the display shows “No - ###” (### stands for the serial number). Tap the CAL 

button to change the serial number from 1-100. Press the MODE button when you are finished.

4.3. Commands

The following commands can be sent to the balance:

Command

Function

Example

serial no. (hex) + s 

Request weight data     

example: 01h s

hex: 01:73

serial no. (hex) + v 

Enter Calibration Mode

example: 01h v

hex: 01:76

serial no. (hex) + c 

Enter Counting Mode

example: 01h c

hex: 01:63

serial no. (hex) + u 

Change weighing unit

example: 01h u

hex: 01:75

serial no. (hex) + t 

Tare the balance (zero)

example: 01h t

hex: 01:74

NOTE: Hex values are notated by a lowercase h after the hex pair (ie. 01h reprents the hex pair 01)

Summary of Contents for KG-10

Page 1: ...INSTRUCTION MANUAL KG Series High Capacity Precision Balance American Weigh Scales Inc KG 10 KG 20 2010 American Weigh Scales Inc Doc Rev 2 1...

Page 2: ...s p 5 2 4 Turning on the Balance and Self testing p 5 3 OPERATION 3 1 Sample Weighing p 6 3 2 Weighing with a Container p 6 3 3 Unit Switching p 6 3 4 Parts Counting p 6 3 5 Percent Weighing p 6 3 6 C...

Page 3: ...1 KG Series Balance Structure 1 Weighing Pan 2 Weight Display 3 Bubble Level 4 ON OFF Button 5 UNIT Button 6 TARE Button 7 MODE Button 8 CAL Button 9 Leveling Foot 10 Battery Compartment Figure 1 3 4...

Page 4: ...Calibration Weight 5000g 10 000g Linear Calibration Weights 5000g 10000g 10 000g 20 000g Response Time 2 3 sec Pan Size 12 x 9 inch Dimensions 12 x 12 5 x 7 inch Net Weight 6 4 lb Power Requirements...

Page 5: ...he balance on a level surface free of vibration Carefully set the scale platform into place You may want to keep the original box and packing materials for storing the balance when not in use If the b...

Page 6: ...ts with each press of the button Notes The balance will default to the last unit used when turned on the next time 3 4 Parts Counting With items of uniform weight ie all the pieces are of equal weight...

Page 7: ...splay 5 Place the calibration weight onto the center of the platform and wait for the display to stop flashing 6 You may now remove the weight The scale will return to normal weighing mode 3 6 Linear...

Page 8: ...r demonstration purposes only To install simply copy the program files to your computer and launch Weight Display exe By default the program listens for weight data on serial port1 4 2 Serial Number W...

Page 9: ...hing platform Avoid dropping your scale The warranty does not cover damage due to rough treatment or overload Keep the weighing pan clean using a soft cloth Do not use harsh chemicals 4 5 Troubleshoot...

Reviews: