![AWS AWS-3000 User Manual Download Page 6](http://html.mh-extra.com/html/aws/aws-3000/aws-3000_user-manual_3035588006.webp)
OPERATION 5
RS-232
The AWS-3000 display can be connected to a printer, computer or data
collector via its RS232 interface. Every time a reading is accepted into
memory, a peak is cleared, or data is transmitted via the print data menu(s),
it is transmitted via the RS-232 port. To download the readings, go to the
DATA MENU.
SPECIAL NOTE: Model AWS-3000 PTEST is not equipped with memory, so the
operation of the RS-232 will vary from the standard unit. The protocol,
however, remains the same.
RS232 Transfer Protocol
Protocol
Value
Cable
9 pin to mini-plug.
Baud
9600
Parity
Flow
None
None
Bits
8
S Bit
1
RS232 Datastream Format
mmmbsdddddbuuuuucl,
where:
m
Memory Location
s
Sign (space or -)
c
Carriage Return
d
Data with Decimal Point
l
Line Feed
u
Units
b
Blank
RS232 Cable Pinouts
Pin #
Description
Pin #
Description
1 Unused
6 Unused
2
Transmit
7
Unused
3 Receive
8 Unused
4
Unused
9
Unused
5 Ground
AWS-3000 Guide