Feed Conductivity - Product Conductivity
Rejection (%) =
Feed Conductivity
Since the system conductivity monitor displays the two in a 10 to 1 ratio, the approximate rejection is
available at a glance. If the needle on the meter does not move when the display is shifted from feed
to product it indicates the rejection is 90%. If the needle when indicating product is only half the
value of the feed, then rejection is 95%. If the needle when indicating product is only one fourth of
the value of the feed, then rejection is 97.5%. This is illustrated below for several different feed
waters.