2. Turn on the AVP-AV-1-B. The main menu will appear on the LCD.
3. To set the AVP-AV-1-B to trigger from the internal clock at a PRF of 10 kHz:
a) The arrow pointer should be pointing at the frequency menu item. If it is not,
press the MOVE button until it is.
b) Press the CHANGE button. The frequency submenu will appear. Rotate the
ADJUST knob until the frequency is set at 10 kHz.
c) The arrow pointer should be pointing at the “Internal” choice. If it is not, press
MOVE until it is.
d) Press CHANGE to return to the main menu.
4. To set the delay to 100 ns:
a) Press the MOVE button until the arrow pointer is pointing at the delay menu
item.
b) Press the CHANGE button. The delay submenu will appear. Rotate the
ADJUST knob until the delay is set at 100 ns.
c) The arrow pointer should be pointing at the “Normal” choice. If it is not, press
MOVE until it is.
d) Press CHANGE to return to the main menu.
5. To set the pulse width to 3 ns:
a) Press the MOVE button until the arrow pointer is pointing at the pulse width
menu item.
b) Press the CHANGE button. The pulse width submenu will appear. Rotate the
ADJUST knob until the pulse width is set at 3 ns.
c) The arrow pointer should be pointing at the “Normal” choice. If it is not, press
MOVE until it is.
d) Press CHANGE to return to the main menu.
6. At this point, nothing should appear on the oscilloscope.
7. To enable the output:
25
Summary of Contents for AVP-AV-1-B
Page 32: ...WIRING DIAGRAMS WIRING OF AC POWER...
Page 33: ...PCB 158R5 LOW VOLTAGE POWER SUPPLY...
Page 34: ...PCB 235C HIGH VOLTAGE DC POWER SUPPLY...
Page 35: ...PCB 231A INTERNAL OFFSET GENERATOR...
Page 36: ...PCB 216F PW PRF TWEAK...
Page 37: ...PCB 151D RANGE POLARITY RELAY BOARD...
Page 38: ...PCB 104G KEYPAD DISPLAY BOARD...
Page 39: ...MAIN WIRING...
Page 40: ...PERFORMANCE CHECKSHEET 40...