The next figure illustrates the relationship between the signals when an external TTL-
level trigger is used:
Figure C
As before, if the delay is negative, the order of the SYNC and OUT pulses is reversed.
In general, the delay, pulse width, and frequency (when in the internal mode), of the
OUT pulse can be varied with front panel controls or via the GPIB or RS-232 computer
interfaces.
TRIGGER MODES
This instrument has four trigger modes:
Internal Trigger: the instrument controls the trigger frequency, and generates the
clock internally.
External Trigger: the instrument is triggered by an external TTL-level clock on the
back-panel TRIG connector.
Manual Trigger: the instrument is triggered by the front-panel “SINGLE PULSE”
pushbutton.
Hold Trigger: the instrument is set to not trigger at all.
These modes can be selected using the front panel trigger menu, or by using the
appropriate programming commands. (See the “Programming Manual for -B
Instruments” for more details.)
20
SYNC OUT
100 ns, FIXED
MAIN OUTPUT
PULSE WIDTH
DELAY > 0
AMPLITUDE,
VARIABLE
3V, FIXED
> 50 ns
TTL LEVELS
(0V and 3V-5V)
TRIG
(external input)
PROPAGATION DELAY (FIXED)
Summary of Contents for AVP-AV-1-B
Page 32: ...WIRING DIAGRAMS WIRING OF AC POWER...
Page 33: ...PCB 158R5 LOW VOLTAGE POWER SUPPLY...
Page 34: ...PCB 235C HIGH VOLTAGE DC POWER SUPPLY...
Page 35: ...PCB 231A INTERNAL OFFSET GENERATOR...
Page 36: ...PCB 216F PW PRF TWEAK...
Page 37: ...PCB 151D RANGE POLARITY RELAY BOARD...
Page 38: ...PCB 104G KEYPAD DISPLAY BOARD...
Page 39: ...MAIN WIRING...
Page 40: ...PERFORMANCE CHECKSHEET 40...