SPECIFICATIONS
Model
1
:
AVMP-4-B
Amplitude
2
:
1 - 10 Volts
Pulse width (FWHM):
10 ns - 10 us
Maximum PRF:
1 MHz
Maximum duty cycle:
10%
Rise time (20%-80%):
≤ 200 ps
Fall time (80%-20%):
≤ 200 ps
Required load impedance:
50 Ohms
6
Polarity
3
:
Positive or negative or both (specify)
GPIB and RS-232 control
1
:
Standard on -B units.
LabView Drivers:
http://www.avtechpulse.com/labview
for availability and downloads
Ethernet port, for remote control
using VXI-11.3, ssh, telnet, & web:
Included. Recommended as a modern alternative to GPIB / RS-232.
See
http://www.avtechpulse.com/options/vxi
for details.
Settings resolution:
The resolution of the timing parameters (pulse width, delay, period) varies,
but is always better than 0.15% of (|set value| + 20 ns).
The amplitude resolution is < 0.1% of the maximum amplitude.
Settings accuracy:
Amplitude: Typically ± (3% of setting) ± (2% of maximum).
Delay, Period: Typically ± (3% of setting ) ± (5 ns)
Pulse width: Typically ± (3% of setting ) ± (2 ns), at maximum amplitude.
As the amplitude is reduced, the pulse width may shift ± 5 ns.
For high-accuracy applications requiring
traceable calibration, verify the output with a calibrated oscilloscope
7
.
Propagation delay:
≤ 200 ns (Ext trig in to pulse out)
Jitter:
± 35ps ± 0.015% of sync delay
Trigger modes:
Internal trigger, external trigger (TTL level pulse, > 10 ns, 1 kΩ input impedance),
front-panel “Single Pulse” pushbutton, or single pulse trigger via computer command.
Variable delay:
Sync out to main out: 0 to 1.0 seconds, for all trigger modes (including external
trigger).
Sync output:
> +3 Volts, > 50 ns, will drive 50 Ohm loads
Monitor output option
5
:
Provides a 20 dB attenuated coincident replica of main output
Connectors:
Out, Monitor: SMA, Trig, Sync, Gate: BNC
Power requirements:
100 - 240 Volts, 50 - 60 Hz
Dimensions: (H x W x D)
100 mm × 430 mm × 375 mm (3.9” × 17” × 14.8”)
Chassis material:
Aluminum frame & handles, blue vinyl on aluminum covers
Temperature range:
+5°C to +40°C
1) -B suffix indicates IEEE-488.2 GPIB and RS-232 control of amplitude, pulse width, PRF and delay (See
http://www.avtechpulse.com/gpib
).
2) For operation at amplitudes of less than 20% of full-scale, best results will be obtained by setting the amplitude near full-scale and using external
attenuators on the output.
3) Indicate desired polarity by suffixing model number with -P or -N (i.e. positive or negative) or -PN for dual polarity option.
4) The -OS option adds an internal bias tee circuit to the output, allowing an externally-generated DC offset (±50V/250mA max) to be
added the output signal. The -OT option includes the -OS function, and adds the ability to generate a ±5V/100mA offset internally
(controlled from the front panel, or by computer command on -B units).
5) Add -M to model number for monitor option.
6) A 50 Ohm load is required. Other loads may damage the instrument. Consult Avtech ([email protected]) if you need to drive other load
impedances.
7) These instruments are provided with a basic calibration checksheet, showing a selection of measured output parameters. These measurements are
performed with equipment that is calibrated on a regular basis by a third-party ISO/IEC 17025:2005 accredited calibration laboratory. However,
Avtech itself does not claim any accreditation. For applications requiring traceable performance, use a calibrated measurement system rather than
relying on the accuracy of the pulse generator settings.
6
Summary of Contents for AVMP-4-B
Page 27: ...This completes the operational check 27...
Page 35: ...PCB 158R4 LOW VOLTAGE POWER SUPPLY...
Page 39: ...PCB 151D RANGE POLARITY RELAY BOARD...
Page 42: ...PERFORMANCE CHECKSHEET...