REAR PANEL CONTROLS (UNITS WITHOUT -FPD OPTION)
Note: some connectors may be in different positions than shown above, depending on
the exact combination of options ordered.
1. AC POWER INPUT
. An IEC-320 C14 three-pronged recessed male socket is
provided on the back panel for AC power connection to the instrument. One end of
the detachable power cord that is supplied with the instrument plugs into this socket.
2. AC FUSE DRAWER
. The two fuses that protect the AC input are located in this
drawer. Please see the “FUSES” section of this manual for more information.
3. DC FUSES
. These two fuses protect the internal DC power supplies. Please see the
“FUSES” sections of this manual for more information.
4. GATE
. This TTL-level (0 and +5V) logic input can be used to gate the triggering of
the instrument. This input can be either active high or active low, depending on the
front panel settings or programming commands. (The instrument triggers normally
when this input is unconnected). When set to active high mode, this input is pulled-
down to ground by a 1 kΩ resistor. When set to active low mode, this input is pulled-
up to +5V by a 1 kΩ resistor.
5. TRIG
. This TTL-level (0 and +5V) logic input can be used to trigger the instrument, if
the instrument is set to triggering externally. The instrument triggers on the rising
edge of this input. The input impedance of this input is 1 kΩ. (Depending on the
21
4
9
3
5
12
6
7
10
11
2
1
8
GATE
RS-232
GPIB
HV
PULSE
OUT
LOGIC
OUT
TRIG
VCC2
OUT
A
B
Summary of Contents for AVRQ-4-B
Page 41: ...41...
Page 51: ...Bottom side 51...
Page 67: ...PCB 158R5 LOW VOLTAGE POWER SUPPLY...
Page 68: ...PCB 104G KEYPAD DISPLAY BOARD...
Page 69: ...PCB 228C OP AMPS AHV AND XHV UNITS...
Page 71: ...DUT WIRING ON STANDARD DAUGHTERBOARD PCB 267D...
Page 72: ...DUT WIRING ON SO 8 DAUGHTERBOARD PCB 314A...
Page 73: ...DUT WIRING ON WDIP16 DAUGHTERBOARD PCB 325A...
Page 75: ...PERFORMANCE CHECK SHEET 75...