INSTALLATION
VISUAL CHECK
After unpacking the instrument, examine to ensure that it has not been damaged in
shipment. Visually inspect all connectors, knobs, liquid crystal displays (LCDs), and the
handles. Confirm that a power cord, a GPIB cable, and two instrumentation manuals
(this manual and the “Programming Manual for -B Instruments”) are with the
instrument. If the instrument has been damaged, file a claim immediately with the
company that transported the instrument.
The following additional items should be with the instrument (for standard models):
1. Eight assembled PCB 267D daughterboards
2. Eight A23-type 12V batteries
Additional items may be included, depending on the options ordered.
POWER RATINGS
This instrument is intended to operate from 100 - 240 V, 50 - 60 Hz.
The maximum power consumption is 57 Watts. Please see the “FUSES” section for
information about the appropriate AC and DC fuses.
This instrument is an “Installation Category II” instrument, intended for operation from a
normal single-phase supply.
CONNECTION TO THE POWER SUPPLY
An IEC-320 three-pronged recessed male socket is provided on the back panel for AC
power connection to the instrument. One end of the detachable power cord that is
supplied with the instrument plugs into this socket. The other end of the detachable
power cord plugs into the local mains supply. Use only the cable supplied with the
instrument. The mains supply must be earthed, and the cord used to connect the
instrument to the mains supply must provide an earth connection. (The supplied cord
does this.)
Warning: Failure to use a grounded outlet may result in injury or death due to
electric shock. This product uses a power cord with a ground connection. It must be
connected to a properly grounded outlet. The instrument chassis is connected to the
ground wire in the power cord.
The table below describes the power cord that is normally supplied with this instrument,
depending on the destination region:
13
Summary of Contents for AVRQ-4-B
Page 41: ...41...
Page 51: ...Bottom side 51...
Page 67: ...PCB 158R5 LOW VOLTAGE POWER SUPPLY...
Page 68: ...PCB 104G KEYPAD DISPLAY BOARD...
Page 69: ...PCB 228C OP AMPS AHV AND XHV UNITS...
Page 71: ...DUT WIRING ON STANDARD DAUGHTERBOARD PCB 267D...
Page 72: ...DUT WIRING ON SO 8 DAUGHTERBOARD PCB 314A...
Page 73: ...DUT WIRING ON WDIP16 DAUGHTERBOARD PCB 325A...
Page 75: ...PERFORMANCE CHECK SHEET 75...