ZM400 Series Indicators Service Manual
61
6.4
System
Use this to print the indicator configuration information. You
pick the port to use for printing in the following steps. See the
extensive list in Chapter
Choices are
Port 1
,
Port 2
and
uSb
(text file).
rESEt
Under this item you can choose to reset the
APP.cnFg
(reset
configuration settings back to how the application had it set
when it was last downloaded),
conFig
(factory default
settings) and
cAL
(factory default calibration) memory or
choose
ALL
to reset both config and cal factory defaults at
once.
6.4.8 Archive
ArchiVE
This stands for electronic archive. In other regions it is known as an alibi
memory or tally roll. These will be transmitted via configured
communications ports. These reports can be used to confirm a weighment
if it is ever in dispute. Only records that are qualified as a LEGAL record,
per site motion criteria, are considered as entries to this log.
EnAbLE
Choose
on
to enable a PC query. If enabled, a
menu
item appears to the right of
EnAbLE
.
You can choose to print the archive from Port 1, 2 or to USB.
(USB device must be installed)
If USB is selected, a USB flash drive must be installed to create the text file of the
indicator configuration.
Archive
Enable
Off
On
Port 1
Port 2
USB
Appears if
Archive enabled
The indicator has memory capacity to store approximately 5,000 transactions.
The Archive report uses a rolling memory configuration (FIFO) so the oldest
transaction will be written over first when all the memory slots are filled.