10
Count
Count
Choose this to display the raw, digital counts from the
Quartzell
TM
. The raw count should be approximately 90,000. If
tension and compression crystal counts are present and stable,
the QDT cell is functioning properly.
SERIAL
Choose this to perform a serial loopback test. You must connect
the transmit and receive lines together for this test to work.
Display will show pass or fail.
dISPLAY
Display
Choose this to display the model number, software version part
number and revision and to perform a display segment test.
bUttonS
Buttons
Choose this to test the keys of the front panel. Each key press is
spelled out in the display to show it is functioning properly. The
COUNT/WEIGHT
key will return the display to
bUttonS
.
HI RESA
High resolution with AZT enabled
Choose this to display weight data with 1,000,000 count resolu-
tion. AZT is enabled during this test. Press the
SAMPLE
or
COUNT/WEIGHT
key to stop the test.
HI RESn
High resolution with no AZT
Choose this to display weight data with 1,000,000 count resolu-
tion. AZT is disabled during this test. Press the
SAMPLE
or
COUNT/WEIGHT
key to stop the test.
Summary of Contents for PC-805
Page 1: ...PC 805 Parts Counter Service Manual...
Page 2: ...2...
Page 20: ...20...
Page 24: ...PC 805 COUNTING SCALE SYSTEM WIRING BLOCK DIAGRAM 24...
Page 25: ...PC 805 COUNTING SCALE MAIN PC BOARD DISPLAY BD KEYPAD OVERLAY 25...
Page 26: ...PC 805 COUNTING SCALE BATTERY INSTALLATION Battery Kit P N 29676 0010 26...
Page 27: ...PC 805 COUNTING SCALE SCALES AND BASES DIMENSIONAL OUTLINE FOR 12 x 14 BASE 27...