3.2 Gap/Black Mark Calibration
Self-test
While calibrate the gap/black mark sensor, printer will measure the label length, print
the internal configuration (self-test).
To calibrate gap or black mark sensor, depends on
the sensor setting in the last print job.
Please follow the steps below to calibrate the sensor.
1.Turn off the power switch.
2. Hold on the button then turn on the power switch.
3. Release the button when LED becomes
amber
and blinking. (Any amber will do
during the 5 blinks).
The LED color will be changed as following order.
Amber
red (5 blinks)
amber (5 blinks)
green (5 blinks)
green/amber (5 blinks)
red/amber (5 blinks)
solid green
4. It calibrates the sensor and measures the label length and prints internal settings.