FTK100 FIELD TEST KIT
OPERATING MANUAL
Document-P/N:
FTK100-OM
© Aventech Research Inc. 2013. All Rights Reserved.
Solution status is broadcasted by GPS to indicate the status and errors found.
Following table describes the status and descriptions.
Table 2: Solution Status and Description
Solution
Status
Description
0
Solution computed
1
Insufficient observations
2
No convergence
3
Singularity at parameters matrix
4
Covariance trace exceeds maximum(trace > 1000m)
5
Test distance exceeded(maximum of 3 rejections if
distance>10km)
6
Not yet converged from cold start
7
Height or velocity limits exceeded
8
Variance exceeds limits
9
Residuals are too large
10
Delta position is too large
11
Negative variance
12
Reserved
13
Large residuals make position unreliable
14-17
INS solution status values
18
When a FIX POSITION command is entered, the receiver
computes its own position and determines if the fixed position is
valid
19
The fixed position, entered using the FIX POSITION command,
is not valid
20
Position type is unauthorized - HP or XP on a receiver not
authorized for it
3.1.7 Thermistor
In this test mode, the temperature readings from the thermistor and built-in temperature
sensor in the RH sensor will be compared. The difference between these two
temperatures and average temperature will also be displayed with comments (if any).