alarm mute button
Control and Monitoring System
Temperature Control System:
This highly accurate electronic control, located in the top right of the control panel (pictured below),
incorporates a temperature probe and thermostat.
A temperature probe measures the air temperature in the cabinet.
The electronic controller is programmed to the required temperature parameters (+2 to +8°C).
The controller cycles the refrigeration system to maintain the air temperature inside the
cabinet, within the required parameters.
Should the temperature of the cabinet fall or rise outside the set parameters, the controller will trigger
both an audible and visual alarm (the digital display will flash –“
LO
” for below set parameter or “
HI
”
for above set parameter).
A suitable delay (5-10 min.) is programmed into the system to prevent unnecessary false alarms (such
as door being open for extended periods).
If an alarm condition occurs, press the
alarm mute button
to silence the audible alarm.
The electronic controller is pre-programmed, during the manufacturing process.
The only button the operator is required to press is the
set/mute
button. If the cabinet alarms, you can
press this to silence it. Any reprogramming of the controller must be done by a person approved by
Avem Quirks Australia.
When the cabinet is first powered on, if it does not reach a temperature below +8 °C within 10 minutes,
it will alarm. This is a normal function. All the operator is required to do is press the
set/mute
button
to silence the alarm and restart the cooling process
If an alarm condition has occurred, please allow 10-15 minutes for recovery.
Should you experience any issues with the Medisafe refrigerator maintaining temperature within the set
parameters, please contact Avem Quirks Australia.
.