background image

DIGITAL TRANSDUCER TESTER & PC-INTERFACE (AVA-03) 

14-FEB-2022 

PAGE 1 OF 

12

 
 

© Avanti Elektronik Limited  -  (44) 01224-583132  -  www.avanti.clara.net  -  [email protected]                  dqdtst_32.docx 
 

 

1. GENERAL 

 
The  Avanti  I²C  Transducer  Tester  and PC-Interface (AVA-03) is a standalone device for 
accessing pressure/temperature transducers, which are equipped with a two-wire serial I²C 
interface.  It  was  designed  to  be  compatible  with  Quartzdyne's  range  of  digital  quartz 
pressure  transducers,  including  their  V4.02  enhancements  (checksum  etc).  Up  to  4 
transducers can be connected simultaneously and their readings are displayed on a backlit 
LCD.  There  is  also  an  RS232  port  provided,  which  allows  to  access  readings  and 
calibration coefficients from a PC. 
 

2. HARDWARE DESCRIPTION 

 
The AVA-03 I²C Transducer Tester is housed in an aluminium enclosure measuring 168 x 
103 x 56mm. 
 

Transducer Socket Pinout 

Pin-1 

A1 

Pin-2 

SCL 

Pin-3 

GND 

Pin-4 

SDA 

Pin-5 

+5V 

Pin-6 

A2 

Pin-7/8/9 

NC 

 

 

RS-232 Socket Pinout 

Pin-2 

RX (of PC) 

Pin-3 

TX (of PC) 

Pin-5 

GND 

Pin-
1/4/6/7/8/9 

NC 

 
There are four 9-way D-type sockets (A...D) for connecting up to four I²C transducers. The 
connectors  carry  +5V,  GND,  SCL,  SDA  and  two  address  lines  A2  and  A1.  The  address 
lines are either open (1) or grounded (0) and are coded differently at each socket: The first 
socket  sets  A2=0/A1=0,  the  second  socket  sets  A2=0/A1=1,  the  third  socket  sets 
A2=1/A1=0 and the fourth socket sets A2=1/A1=1. This means that different I²C addresses 
are allocated automatically for each socket, provided the address pins are wired through to 
the transducers, see the examples shown in fig. 1 and fig. 2 below. 
 
 

 

Figure 1: 6-wire connection, I²C address coded by the AVA-03 socket. 

Summary of Contents for AVA-03

Page 1: ...ARE DESCRIPTION The AVA 03 I C Transducer Tester is housed in an aluminium enclosure measuring 168 x 103 x 56mm Transducer Socket Pinout Pin 1 A1 Pin 2 SCL Pin 3 GND Pin 4 SDA Pin 5 5V Pin 6 A2 Pin 7...

Page 2: ...be higher than the battery voltage in order to prevent the battery from inadvertently sourcing current Current consumption is approximately 5mA when the backlight is off and approximately 125mA when...

Page 3: ...0B61 A RAW T 0x01C71C72 3 Raw Values PF 20000 000 Hz A TF 50000 000 Hz 4 Equivalent Frequencies P 11111 222 psi A T 111 222 degC 5 Calculated P T Values A2A1 01 XDUCER B SN 000456 14 02 2022 6 Status...

Page 4: ...ansducers together see paragraph 6 2 12 BAR DEGF DUAL P Absolute pressure is displayed in bar and temperature in F Initial screen shows pressure values of 2 transducers together see paragraph 6 2 Once...

Page 5: ...psi B 10 Dual pressure screen again P1 is allocated to the transducer with the lowest I C address and P2 to the transducer with the second lowest I C address If more than two transducers are connecte...

Page 6: ...er Tester to output a list of all implemented RS232 commands plus the FPGA identifier and status of all connected transducers GET A RAW PRESSURE READING COUNTS PC SENDS TESTER RESPONDS DESCRIPTION BYT...

Page 7: ...Echo 1 A CR 0x0D Carriage return 1 CR 0x20 Space 1 0 000 99999 999 Pressure in psi or bar as a 5 3 floating point number 9 1234 567 CR 0x0D Carriage return 1 CR LF 0x0A Line Feed 1 LF GET A CALCULATED...

Page 8: ...Read EEPROM request 1 R R Echo 1 R A D Transducer address 1 A A D Echo 1 A 0000 1FFF Start address 4 0100 0000 1FFF Echo 4 0100 00 FF Number of bytes to read 00 256 2 0A 00 FF Echo 2 0A CR 0x0D Carria...

Page 9: ...K or NO Result of operation 2 OK 0x20 Space 1 0000 FFFF Checksum over all data bytes 4 0171 CR 0x0D Carriage return 1 CR LF 0x0A Line feed 1 LF SET SERIAL NUMBER AND CALIBRATION DATE PC SENDS TESTER R...

Page 10: ...zdyne in Jan 2010 START CONTINUOUS OUTPUT MODE PC SENDS TESTER RESPONDS DESCRIPTION BYTES EXAMPLE CM Continuous Mode request 2 CM CM Echo 2 CM CR 0x0D Carriage return 1 CR CR 0x0D Echo 1 CR LF 0x0A Li...

Page 11: ...charting functions are required there is a low cost 3rd party software available that runs on Windows PCs see www windmill co uk Windmill 7 data logging software includes a user configurable parser th...

Page 12: ...e at http www quartzdyne com The following documents are of particular interest DigitalTransSpec pdf Digital Quartz Pressure Transducer Specifications DigitalTransProg pdf Digital Transducer Programmi...