APT3100
Smart Pressure Transmitter
Operation Manual M3100-EO1G
19
Duon System Co.,Ltd.
(1) Moving between menus : Zero
(2) Enter or moving to sub menu : Span
(3) Moving to top menu : Zero+Span
y
Put the button for 3 seconds to execute each function. After 3 seconds to put the ZERO+SPAN button,
LCD Display will be changed from Menu to Trim. To see next menu, put the Zero Button for 3 seconds
and then Setup Menu will be displayed. Use Zero button to move down to the next directory.
Put the Span button to select the displayed menu. The same procedure will be applicable for the sub
menus.
Caution : After about 30 seconds without any action, the button function will be finished.
y
How to put numerical value
①
Functions needed to put numerical value : 12 Zero Adjustment, 22 Change Upper Range Value,
23 Change Lower Range Value, 24 Damping Second
②
How to put numerical value : Firstly, select a increasing rate(10
n
) and then, change the each
decimal value to increase or decrease as wanted.
③
For example, to select 3810 : Select increasing rate as 1000 -> Increase 3 times as 1000 ->
Select increasing rate as 100 -> Increase 8 times as 100 -> Select increasing rate as 10 ->
Increase 1 time as 10
④
To select the increase / decrease steps : SelInc Message will be displayed on the bottom of LCD.
Select step to put the Zero button : The decimal value will be changed whenever put Zero button.
After set the wanted step, put the span button to execute the item no. 2
⑤
How to set the required value using Zero, Span buttons. : VALUE Message will be displayed on
the bottom of LCD window.
When put the Zero button, the step will be increased as the set on the item 1.
When put the Span button, the step will be decreased as the set on the item 1.
After setting, execute again the item 1 to put Zero+Span button.
⑥
Set the final value to repeat
④
, and
⑤
After setting the final value, finish the procedure to put the Zero+Span button while executing