Autonics KT-502H Quick Start Manual Download Page 3

KT-502H

G-4

Fid dev output 20.000

mA equal to reference 

meter ?

1. Yes
2. No

ABORT ENTER

NOTE-Loop may be

returned to automatic 

control

ABORT

O K

Device Disconnected

RETRY

QUIT

1. Offline

2. Online

3. Frequency Device

4. Utility

Setting fid dev.

output to 20mA

ABORT

O K

Enter meter Value

20.000

HELP

DEL

ABORT ENTER

Diag/Service

1. Test device

2. Loop test

3. Calibration

4. D/A trim

HELP

SAVE

HOME

 Press the 

 (F4) key. 

 Press the 

 (F3) key. 

 Press the 

 (F3) key.

  Press the   (F3) key to complete the adjustment. 

 If output display value is correct, select 

'1. Yes' 

and press 

the 

 (F4) key. If not, select 

'2. No' 

and press the 

 (F4) key and re-set the display value.

 Press the 

 (F4) key. 

 Press the 

 (F4) key to set 20 mA display value.

Current Trim Adjustment

1. Device Setup

2. PV

3. PV Ao

4. PV LRV

5. URV   

1. Process Variables

2. Diag/Service

3. Basic Setup

4. Detailed Setup

5. Review  

1. Test device

2. Loop test

3. Calibration

4. D/A trim

Connect reference

meter

ABORT

O K

Fid dev output 4.000

mA equal to reference 

meter ?

1. Yes
2. No

ABORT ENTER

Enter meter Value

4.000

HELP

DEL

ABORT ENTER

Setting fid dev

output to 4mA

ABORT

O K

WARN-Loop should be

removed from

automatic control

ABORT

O K

 Select the 

'1. Device Setup' 

by 

,   

keys and press the 

       key. 

Connect a HART communicator and adjust current trim as below by a HART communicator.

 Select the 

'2. Diag/Service' 

by 

,   

keys and press the   

key. 

 Select the 

'4. D/A trim' 

by 

,   

 keys and press the   key. 

 Press the 

 (F4) key. 

 Press the 

 (F4) key. 

 Press the 

 (F4) key. 

 Press the 

 (F4) key to set 4 mA display value.

 If output display value is correct, select 

'1. Yes' 

and press 

the 

 (F4) key. If not, select 

'2. No' 

and press the 

 (F4) key and re-set the display value. 

Ex)If output display value is 3.89mA, select 3.89 and press  

the 

 (F4) key. 

Summary of Contents for KT-502H

Page 1: ...100 100mV 100 500mV 100 2000mV Output DC4 20mA 2 wire Accuracy 0 3 Alarm Below 3 8mA Over 20 5mA Sensor break 3 6mA Load Max V power supply 7 5V 0 22A Galvanic insulation 2kVAC input output Environ ment Ambient temperature 20 to 70 storage 20 to 80 Ambient humidity 0 to 85 RH storage 0 to 85 RH Explosion class Ex d IIC T6 Protection structure IP67 IEC standard Material Body Aluminum AlDc 8S Cover ...

Page 2: ... Ω 0 to400Ω 0 to2000Ω Thermocouple B PtRh30 PtRh6 0 to 1820 32 to 3308 E NiCr CuNi 270 to 1000 454 to 1832 J Fe CuNi 210 to 1200 346 to 2192 K NiCr Ni 270 to 1372 454 to 2501 6 N NiCrSi NiSi 270 to 1300 454 to 2372 R PtRh13 Pt 50 to 1768 58 to 3214 4 S PtRh10 Pt 50 to 1768 58 to 3214 4 T Cu CuNi 270 to 400 454 to 752 Analog Voltage 10 75mV 100 100mV 100 500mV 100 2000mV Opening cover Dimensions Co...

Page 3: ...o 4 PV LRV 5 URV 1 Process Variables 2 Diag Service 3 Basic Setup 4 Detailed Setup 5 Review 1 Test device 2 Loop test 3 Calibration 4 D A trim Connect reference meter ABORT O K Fid dev output 4 000 mA equal to reference meter 1 Yes 2 No ABORT ENTER Enter meter Value 4 000 HELP DEL ABORT ENTER Setting fid dev output to 4mA ABORT O K WARN Loop should be removed from automatic control ABORT O K Selec...

Page 4: ...supply or cut off the power Switch or circuit breaker should be installed nearby users for convenient control Do not use this unit near the high frequency instruments high frequency welding machine sewing machine large capacity SCR controller Installation environment Indoor Outdoor Altitude max 2 000m Pollution degree 2 Installation category Ⅱ Use the verified explosion proof electric connection c...