SAVe II+ OPERATOR’S MANUAL > Page 24
Table 3: Adult Height Preset Values
Height
Rate
(BPM)
Tidal volume
(ml)
Minute
Ventilation (LPM)
PIP Limit
(cmH2O)
PEEP
(cmH2O)
4’ 3” (129 cm)
20
250
5.0
30
0
4’ 6” (137 cm)
21
250
5.3
30
0
4’ 9” (145 cm)
21
260
5.5
30
0
5’ 0” (152 cm)
20
300
6.0
30
0
5’ 3” (160 cm)
18
340
6.1
30
0
5’ 6” (168 cm)
16
380
6.1
30
0
5’ 9” (175 cm)
15
420
6.3
30
0
6’ 0” (183 cm)
14
470
6.6
30
0
6’ 3” (191 cm)
13
510
6.6
30
0
S
TEP
5:
V
ERIFY
D
ISCONNECT
A
LARM
PROCEDURE
PASS CRITERIA
Simulate a disconnect by
leaving circuit disconnected
from Patient’s airway
- DISCONNECT visual alarm indicator begins blinking
within 2 breaths
- Audible alarm indicator can be clearly heard
- Pump continues to operate normally (cycle)
S
TEP
6:
V
ERIFY
PIP
R
EACHED
A
LARM
PROCEDURE
PASS CRITERIA
Completely block patient
connection port with your
hand
- PIP REACHED visual alarm indicator begins blinking
within 1 breath
- Pump turns off for a few seconds and then turns on briefly
again until PIP limit is reached again
WARNING:
A
LWAYS VERIFY THE
D
ISCONNECT AND
PIP
R
EACH PATIENT ALARMS
prior to connecting the SAVe
II+ to the patient and following any change to device’s configuration such as by adding or
removing an accessory to the Ventilator Breathing System.
D
O NOT USE
if the device fails the checkout.
Step 7: Connect Breathing Circuit to Airway
Connect SAVe II+ breathing circuit to the patient airway. This should cause the audio
disconnect alarm to stop and for the DISCONNECT visual alarm to stop blinking and
stay lit for 30 seconds. The PIP Reached visual alarm should also stop blinking and
stay lit for 30 seconds.
Verify air is being delivered to patient by adequate chest rise.