SAVe II+ OPERATOR’S MANUAL > Page 21
S
TEP
5:
C
HARGE
B
ATTERY
The battery is recharged by connecting the SAVe II+ to external power (100 – 240
VAC, 50 – 60 Hz) via the AC POWER SUPPLY (PN: M42090). The SAVe II+ will
simultaneously run and charge the internal lithium-ion battery. It takes a little over 1
hour to fully charge the SAVe II+. The required time to fully charge increases by 15 –
40% if the device is recharged while ventilating.
Table 2: Battery Capacity
Illuminated LEDs
Usable Battery Capacity
4 LEDs
> 75%
3 LEDs
> 50%
2 LEDs
> 25%
1 LED
> 10%
1 LED (Blinking)
< 10%
T
O CHARGE THE BATTERY
:
1)
Connect AC Power Supply to appropriate power source (See power input
specifications) and the power input port of the SAVe II+
2)
Verify charge indicator light (lightning bolt icon) is illuminated
3)
Monitor charge status using the Battery Level Indicator
WARNING:
-
D
ELIVERED
T
IDAL
V
OLUME MAY DECREASE SIGNIFICANTLY WITHOUT ALARMING IF FINE
PARTICULATES
(
SAND
)
ENTERS THE DEVICE MANIFOLD
.
-
N
EVER USE A WET OR MOIST DEBRIS FILTER
.
The debris filter is not designed to filter
chemical or biological agents and will not protect the patient from contaminated
environments.
-
N
EVER OPERATE THE UNIT WITHOUT A DEBRIS FILTER IN PLACE
.
Immediately take device
out of service if dust, sand, or other debris have entered the internal SAVe II+
system.
-
O
NLY USE DEBRIS FILTERS DESIGNED FOR THE
SAV
E
II+.
Using other debris filters may
impact device performance by either allowing fine particulates into the device
manifold or increasing resistance at the air intake port of the ventilator.
Figure 13: Power Input Port