6 MAINTANENCE 

AND 

SERVICE 

 
Before any service verify that the power has been removed from the 
system 
 

6.1 ABSORBER 

REMOVAL 

The service interval of the absorber of the M600 is 20,000 hours (27.4 
months).

   

 

The M600 absorber has been placed at the front of the unit, behind a 
removable cover, to facilitate replacement.  Absorber replacement 
should take about 15 minutes.  This is ideal for rack and rail mounted 
compressor units. 

 
6.2 HELIUM 

PRESSURE 

ADJUSTMENTS 

The helium charge can be increased or decreased by using the ¼” self-
sealing connector on the front of the unit, this connection is marked 
by a white washer. 
 
Never charge the compressor above a static pressure of 250psi 
(17.2bar)

 

 

 
 
 
 
 
7. TROUBLE 

SHOOTING

 

 

For any trouble or questions installing and operating the M600 unit 
please contact the service department of Austin Scientific at: 

 

 

 
Austin Scientific   
Phone  

(512) 441-6893 x 112 

Fax  

(512) 

443-6665 

Summary of Contents for M600

Page 1: ...L CONNECTION 3 3 1 POWER INPUT 3 3 2 MAIN CABLE AND LEADS 3 3 3 COLD HEAD CONTROLS 3 3 4 AUTO RESTART FUNCTION 4 INSTALATION 4 1 UNPACKING AND INSPECTION 4 2 MOUNTING 4 2 1 OPERATING ANGLE 4 2 2 FLEXLINES AND CONNECTIONS 4 2 3 COLD HEAD CONNECTION 5 OPERATION 5 1 SWITCHING ON 5 2 NORMAL OPERATION 5 3 SAFETY DEVICES 6 MAINTANANCE AND SERVICE 6 1 ABSORBER REMOVAL 6 2 HELIUM PRESSURE ADJUSTMENT 7 TRO...

Page 2: ...t rack or cabinet installation Both high and low side pressure gages Electrical box is a field replaceable unit High and low voltage units available for use at both 50 Hz and 60 Hz Water flow meter provides user feedback to ensure adequate water supply Compressor can be controlled remotely ...

Page 3: ...is for the helium Removing the heat of compression from the helium makes it better able to chill the cold head The two hot fluids are cooled by water The water enters and exits the compressor from the front panel The helium leaving the compressor still has a small amount of oil mixed with it If this oil makes it to the cold head it will freeze and foul the cold head The function of the mist separa...

Page 4: ...Flow diagram of the compressor system ...

Page 5: ...200 60HZ 3 PHASE V 460 230 POWER REQUIREMENTS 50HZ 3 PHASE KW 8 2 7 5 60HZ 3 PHASE KW 8 8 7 8 OPERATING CURRENT WITH WARM COLD HEAD 50HZ A 10 21 60HZ A 10 21 REMOTE CONTOL VOLTAGE VAC 24 24 HELIUM CONNECTORS IN Aeroquip Same STATIC HELIUM PRESSURE 50HZ PSI bar 240 16 5 240 16 5 60HZ PSI bar 240 16 5 240 16 5 458 18 0 532 21 0 441 17 4 ...

Page 6: ...mbient temperature must remain between 40 F and 104 F 5 C and 40 C 3 2 2 COOLANT CONNECTIONS The standard connections on this compressor for the coolant system are 3 8 Swagelok connectors Connections can be modified to meet customer requirements 3 2 3 COOLANT QUALITY Heat exchanger manufacturer recommends 1200 840 micron strainer on water supply and the following water requirements Electrical cond...

Page 7: ...s and cables to meet customer requirements 3 3 4 AUTO RESTART FUNCTION Our unit will have an automatic restart with brownout or blackout in power when power is reapplied to the system A manual reset will be necessary for a unit that stops operating with a fault condition 3 4 MULTI CRYO SYSTEMS 3 4 1 COMPRESSOR CAPACITIES The stock compressor has the ability to supply multiple cold heads Compressor...

Page 8: ...ected contact the carrier and Austin Scientific as soon as possible 4 2 MOUNTING 4 2 1 MAX OPERATING ANGLE The compressor unit can be installed and operate with a maximum tilt of 10 in any direction 4 2 2 FLEXLINES and CONNECTORS The lines and connectors have been equipped with self sealing coupling this maintains the positive seal so no helium can escape during any maintenance These couplings sho...

Page 9: ...rked with a red washer to the outlet of the compressor This can be changed with customer requirement Crossing the connection on the cold head will cause damage to the internals of the cold head With the appropriate cable supplied from Austin Scientific complete the connections form the compressor electronic box to the electrical connection on the cold head This will supply power and logic to the c...

Page 10: ...igh pure quality 99 9999 pure any impurities can cause problems inside the cold head that the compressor is controlling 5 3 SAFETY DEVICES The unit is fitted multiple safety devices An external pressure relief valve A thermal switch on the oil temperature A thermal switch on the helium temperature Fuses for the main power Fuses for the cold head An internal pressure relief valve Electronic Phase m...

Page 11: ...ke about 15 minutes This is ideal for rack and rail mounted compressor units 6 2 HELIUM PRESSURE ADJUSTMENTS The helium charge can be increased or decreased by using the self sealing connector on the front of the unit this connection is marked by a white washer Never charge the compressor above a static pressure of 250psi 17 2bar 7 TROUBLE SHOOTING For any trouble or questions installing and opera...

Page 12: ...eceive special instructions and to obtain a Return Good Authorization RGA number Austin Scientific 4114 Todd Lane P O BOX 18863 Austin TX 78760 512 441 6893 FAX 512 443 6665 Austin Scientific West Coast Service Center 500 Laurelwood Road 9 Santa Clara CA 95054 408 496 0782 FAX 408 469 0785 ...

Page 13: ...TION IN PART OR AS A WHOLE WITHOUT THE WRITTEN PERMISSION OF AUSTIN SCIENTIFIC IS PROHIBITED PROPRIETARY AND CONFIDENTIAL NEXT ASSY USED ON APPLICATION DIMENSIONS ARE IN INCHES TOLERANCES SURFACE FINISH 32RMS OR BETTER ANGULAR MACH 1 BEND 2 TWO PLACE DECIMAL 01 THREE PLACE DECIMAL 005 INTERPRET GEOMETRIC TOLERANCING PER MATERIAL FINISH DRAWN CHECKED ENG APPR MFG APPR Q A COMMENTS DATE NAME AUSTIN ...

Reviews: