AUREL Magna Series Manual Download Page 30

30

Set1

Set2

Output

Alarms

Digit

OK

Max/Min

Reset

Signal  Filtering  /  Averaging

You can adjust the filtering time constant to reduce the effect of noise or instability on your
input signal.

A larger FIL value will give a more stable display, but the response to signal changes will be
slower.

Because your output options, such as analogue output, alarm relays and serial output are all
derived from the displayed value, they will respond at the same rate as the filtered display.

Lockout Switch must be OFF

Press 3 seconds

Set1

Set2

Output

Alarms

Digit

OK

Max/Min

Reset

Display shows  

FIL

.XX.X

  where

XX.X

 is the time constant in seconds.

Use the UP or DOWN button to
increase or decrease this value
 

(Default = 

FIl. 00.0

)

Set1

Set2

Output

Alarms

Digit

OK

Max/Min

Reset

Press to accept

1

2

3

4

Done!

See also Filter Jump setting in the Advanced Menu, if your signal is particularly noisy and
you cannot get sufficient smoothing with this filter.

This  feature  is  available  in  Easy  and  Advanced  Modes

OFF

Circuit board        ON

Set1

Set2

Output

Alarms

Digit

OK

Max/Min

Reset

Display shows  

HFI.

X

  where 

X

 is

the amount of basic filtering.

A

=Low

 b

=Medium

 C

=High

 d

=Most

 

(Default = 

HFI.C

)

Select, then press OK to accept

Select, then press OK to accept

Summary of Contents for Magna Series

Page 1: ...Magna Series 4 digit process display LAUREL Electronics Inc 3183 G Airway Ave Costa Mesa CA 92626 USA Tel 714 434 6131 Fax 714 434 3766 Website www laurels com...

Page 2: ...foregoing warranty shall not apply to defects resulting from 1 Improper or inadequate maintenance by the buyer 2 Unauthorised modification or misuse 3 Operation outside the environmental specification...

Page 3: ...ons Front Buttons 28 Factory defaults 29 CalibrationAudit number 29 Signal Filtering Averaging 30 31 Last Digit rounding 32 Scale Factor adjustment 33 OffsetAdjustment 34 Menu Timeout adjustment 35 Re...

Page 4: ...is equipment must have anti surge T fuses rated at 1A for 230V supply 2A for 110V supply 5A for 48VAC supply or 10A for 11 30VDC Don t touch any circuitry after you have connected the display because...

Page 5: ...shutdown periods you may have to to wait several weeks for an equivalent replacement or we may have no stock at the time you urgently need it You mayalso needto pay extra carriagecharges ifyou wanta f...

Page 6: ...an connect 4 remote wired pushbuttons to the display so that you can make adjustments while the display is mounted in an inaccessible location Power supply options 100 240 VAC 48VAC or 11 30VDC These...

Page 7: ...6 12 4 digit 1004 408 1050 1034 12 6 digit 1494 408 1540 1524 16 4 digit 1322 515 1368 1352 16 6 digit 1974 515 2020 2004 77 mm H mm You can order these displays with the cable glands in the bottom su...

Page 8: ...246 820 8 4 digit 751 290 750 8 6 digit 1073 290 1072 12 4 digit 1051 408 1050 12 6 digit 1541 408 1540 16 4 digit 1369 515 1368 16 6 digit 2021 515 2020 8 8 8 8 8 8 OK dg Xmm 60 mm 80 mm 72 5 mm 60...

Page 9: ...434 452 4 6 digit clock 213 5 195 5 653 671 4 6 digit numeric 213 5 195 5 616 634 6 4 digit 264 246 580 598 6 6 digit 264 246 820 838 8 4 digit 308 290 750 768 8 6 digit 308 290 1072 1090 12 4 digit 4...

Page 10: ...possible locations for the input board output board and power supply board s Their locations depend on the height of digits number of digits brightness of digits and any installed options Because the...

Page 11: ...e remote programming buttons to allow adjustment of the display s settings when the display is inaccessible You can also enable or disable the display s front panel buttons either by a remote contact...

Page 12: ...cancelled Mixing destinations carelessly amongst the twisted pairs can actually worsen noise performance 3 The cable should be routed away from noisy wiring and devices such as power feeds frominverte...

Page 13: ...several displays to several sensors please be sure to use one twisted pair for each display and sensor Do NOT use a wire from one pair for signal positive and a wire from another pair for signal nega...

Page 14: ...Easy orAdvanced mode How to choose menu mode Set1 Set2 Output Alarms Digit OK Max Min Reset Lockout Switch must be OFF Press together briefly Set1 Set2 Output Alarms Digit OK Max Min Reset Press brie...

Page 15: ...ckout Switch must be OFF Press together briefly Set1 Set2 Output Alarms Digit OK Max Min Reset Display shows L Eng Default for Language English or L FrA for Language French 1 2 3 Set1 Set2 Output Alar...

Page 16: ...ss of the UP button will select a newbrightness level There are 7 brightness levels to choose from Default Full brightness Set1 Set2 Output Alarms Digit OK Max Min Reset Press to accept 1 2 3 Done Did...

Page 17: ...each signal level How to choose a calibration method Lockout Switch must be OFF Press 3 seconds Set1 Set2 Output Alarms Digit OK Max Min Reset Press to select dirECt or thEor Display shows CAL then d...

Page 18: ...1 Set2 Output Alarms Digit OK Max Min Reset Press OK Apply the highest calibration signal you can achieve ideally 100 of system capacity You can use less but you will get better accuracy with higher s...

Page 19: ...s OK Max Min Reset Lockout Switch must be OFF Set1 Set2 Output Alarms Digit OK Max Min Reset Press OK Apply the lowest calibration signal you can achieve ideally 0 of system capacity Set1 Set2 Output...

Page 20: ...put Alarms Digit OK Max Min Reset If the display shows Lin O press Set2 button briefly so that the display shows Lin 1 then press OK Set1 Set2 Output Alarms Digit OK Max Min Reset 1 2 3 4 Set1 Set2 Ou...

Page 21: ...button until the display is set as required Press OK when done The display will show Ad Pt Press OK if you want toAdd another calibration point you will repeat the previ ous steps each time increment...

Page 22: ...20mA Set1 Set2 Output Alarms Digit OK Max Min Reset Press to accept Set1 Set2 Output Alarms Digit OK Max Min Reset Display now shows rd Hi briefly You can now specify the maximum reading and decimal...

Page 23: ...the UP or DOWN arrow to increase or decrease each digit s value as re quired eg set 000 0 if your display range is 0 to 100 0 Theoretical Calibration Low end calibration This is when you type in the s...

Page 24: ...then press OK Set1 Set2 Output Alarms Digit OK Max Min Reset 1 2 3 4 The display will show In 00 followed by an editing screen Here you can enter your first calibration input signal level using DIGIT...

Page 25: ...utton and increase or decrease each digit s value using the UP or DOWN button until the display is set as required Press OK when done The display will show Ad Pt Press OK if you want toAdd another cal...

Page 26: ...r readings less than 8 least significant counts The corrected value is stored in non volatile memory so that any drift trend will be re ap plied after you have switched off the meter Lockout Switch mu...

Page 27: ...ion for Contact Closure 1 After you have set CC 1 you will get the prompt CC 2 to allow you to set Contact Closure 2 function and when you have set CC 2 you will get the prompt CC 3 to allow you to se...

Page 28: ...panel Tare Peak Valley and Reset buttons also Or simply link NPN could be opto isolators if you need the logic control lines to be galvanically isolated from the input signal tare Tares display to 0 O...

Page 29: ...Done This feature is available in Easy and Advanced Modes Display shows def n Defaults no Press the DOWN button to change the display to def y Defaults Yes if you want to return to default conditions...

Page 30: ...econds Set1 Set2 Output Alarms Digit OK Max Min Reset Display shows FIL XX X where XX X is the time constant in seconds Use the UP or DOWN button to increase or decrease this value Default FIl 00 0 Se...

Page 31: ...ter is less than 10 per con version Guidance For noisy systems increase the Filter Jump value up to a maximumof 99 Choose a value which gives a good compromise between filtering and response speed For...

Page 32: ...nd up to the nearest 1 no rounding 2 5 10 20 or 50 Lockout Switch must be OFF Press 3 seconds Set1 Set2 Output Alarms Digit OK Max Min Reset Press repeatedly until you see Ld X where X is either 1 2 5...

Page 33: ...ll now read in imperial gallons Changing weight units of measure from kg to pounds You could also use the SCALE to convert your readout from kg to pounds without affecting the calibration Simply set S...

Page 34: ...efault is 0 000 1 2 3 4 Set1 Set2 Output Alarms Digit OK Max Min Reset Use DIGIT button to select each digit in turn UP or DOWN buttons to increase or decrease each digit s value If you want to set a...

Page 35: ...must be OFF Press together briefly Set1 Set2 Output Alarms Digit OK Max Min Reset Press repeatedly until you see dY XX where XX is the delay in seconds Choices are dy 1 0 dy 20 dy 30 dy 60 default 1...

Page 36: ...t2 Output Alarms Digit OK Max Min Reset Press OK button briefly and repeatedly until you see rEV 0 Default or rEV 1 Set1 Set2 Output Alarms Digit OK Max Min Reset Press briefly to toggle 1 2 3 4 Set1...

Page 37: ...a button is pressed Lockout Switch must be OFF Press together briefly Set1 Set2 Output Alarms Digit OK Max Min Reset Press OK button briefly and repeatedly until you see bt 0 or bt 1 or bt 2 or bt 3...

Page 38: ...n 4 Press the Up arrow to change display to def Y and press OK 5 If you want all channels to share a common setting eg calibration do that setting now 6 When you want to do separate settings for each...

Page 39: ...re than this the resistor has probably been overloaded and burned by having a voltage connected across it Please check your field wiring to en sure that no more than 30mA or 1V can be applied to the c...

Page 40: ...the thumb screws on the rear case store them safely and remove the back panel 3 Locate the main option board which will be similar in appearance to the diagram below If a main option board is absent...

Page 41: ...o dispose of it It must not be thrown away with general rubbish If you are unable to dispose of this item locally you may send it to us for recycling Conditions 1 We will only accept items of our manu...

Page 42: ...r current inputs Accuracy 0 05 of range Span tempco 25 ppm Degree Celsius Zero Tempco 30 ppm Degree Celsius 24V Excitation voltage 24VDC nominal rated at 60mA 10V Excitation voltage 10VDC nominal rate...

Page 43: ...board havingswitched capacitor input filter feature Code will auto detect this board If a previous version board is installed the meter will skip switched capaciter filter section of menu The newly re...

Page 44: ...routed separately to power carrying cabling includes relay output wiring All signal cabling shall be screened The screen shall only be terminated to the power earth terminal at the meter end of the c...

Reviews: