Guidance and manufacturer’s declara>on – electromagne>c immunity
The Aura - Nova HD is intended for use in the electromagnetic environment specified below. The customer or
the user of the Aura - Nova HD should assure that it is used in such an environment.
IMMUNITY test
IEC 60601 TEST LEVEL
Compliance level
Electromagne>c environment - guidance
I
MMUNITY
to
proximity fields
from RF wireless
communications
equipment
MHz – Modula>on – Field
Strength
385 - 18 Hz - 27 V/m
450 - 18 Hz - 28 V/m
710 - 217 Hz - 9 V/m
745 - 217 Hz - 9 V/m
780 - 217 Hz - 9 V/m
810 - 18 Hz - 28 V/m
870 - 18 Hz - 28 V/m
930 - 18 Hz - 28 V/m
1720 - 217 Hz - 28 V/m
1845 - 217 Hz - 28 V/m
1970 - 217 Hz - 28 V/m
2450 - 217 Hz - 28 V/m
5240 - 217 Hz - 9 V/m
5500 - 217 Hz - 9 V/m
5785 - 217 Hz - 9 V/m
MHz – Modula>on – Field
Strength
385 - 18 Hz - 27 V/m
450 - 18 Hz - 28 V/m
710 - 217 Hz - 9 V/m
745 - 217 Hz - 9 V/m
780 - 217 Hz - 9 V/m
810 - 18 Hz - 28 V/m
870 - 18 Hz - 28 V/m
930 - 18 Hz - 28 V/m
1720 - 217 Hz - 28 V/m
1845 - 217 Hz - 28 V/m
1970 - 217 Hz - 28 V/m
2450 - 217 Hz - 28 V/m
5240 - 217 Hz - 9 V/m
5500 - 217 Hz - 9 V/m
5785 - 217 Hz - 9 V/m
NOTE: These guidelines may not apply in all situa>ons. Electromagne>c propaga>on is affected by absorp>on and
reflec>on from structures, objects and people.
Portable and mobile RF communica>ons
equipment should be used no closer to any
part of the Aura - Nova HD, including
cables, than the recommended separa>on
distance calculated from the equa>on
applicable to the frequency of the
transmiXer.
Recommended separa>on distance
E = [6/d] √P
d = [6/E] √P
where P is the maximum output power
ra>ng of the transmiXer in waXs (W)
according to the transmiXer manufacturer,
d is the recommended separa>on distance
in meters (m), and E is the field strength in
V/m. Field strengths from fixed RF
transmiXers, as determined by an
electromagne>c site survey, should be less
than the compliance level in each
frequency range. Interference may occur in
the vicinity of equipment marked with the
following symbol:
p