32
Problem
Possible Solution
deterioration.
Reduce patient motion.
Verify that the Infant Oximeter Module is paired
with the Display Unit.
Data update period
has exceeded limit
Reposition the Infant Oximeter Module and
keep the sensor motionless for at least 10
seconds.
Relocate the Infant Oximeter Module at a
different site.
A dash (---) appears
in the vital sign
displays.
Make sure the Infant Oximeter Module is
securely attached to the foot.
Reposition the Infant Oximeter Module and
keep the sensor motionless for at least 10
seconds.
Relocate the Infant Oximeter Module at a
different site.
Turn off the unit, check all connections, and
retry.
Verify that the Infant Oximeter Module is paired
with the Display Unit.
The unit is in Alarm
mode, but no
audible alarms can
be heard.
Wait for two minutes and alarm tones will
automatically re-engage.
Make sure there are no headphones inserted
into the headphone jack of the Display Unit.
Cannot connect the
devices
Turn off both the Infant Oximeter Module and
the Display Unit. Follow instructions in the
“Device Pairing” section on page 16.
Ensure that the Infant Oximeter Module is in
range while being paired (approximately 10
meters (32.8 feet) spherical radius).
If these solutions do not correct the problem, please contact your
distributor, or contact Aulisa Customer Support at 1 (650) 813-1273
(USA).
WARNING!
This system is a precision electronic instrument and
must be repaired by knowledgeable and specially
trained Aulisa personnel only. Do not attempt to open
the case or repair the electronics.